3- 18
85054B
Use, Maintenance, and Care of the Devices
Connections
3. Apply downward force perpendicular to the wrench handle. See
. This applies
torque to the connection through the wrench.
Do not hold the wrench so tightly that you push the handle straight down along its length
rather than pivoting it, otherwise you apply an unknown amount of torque.
4. Tighten the connection just to the torque wrench break point. The wrench handle gives way
at its internal pivot point. See
. Do not tighten the connection further.
CAUTION
You don’t have to fully break the handle of the torque wrench to reach the
specified torque; doing so can cause the handle to kick back and loosen the
connection. Any give at all in the handle is sufficient torque.
Do not pivot the wrench handle on your thumb or other fingers, otherwise you apply an
unknown amount of torque to the connection when the wrench reaches its break point.
Do not twist the head of the wrench relative to the outer conductor mating plane. If you do,
you apply more than the recommended torque.
Connecting the Sliding Load
Use this procedure to connect the sliding load to a test port or a type-N cable connector.
NOTE
The sliding load uses a plastic centering bead to support its center conductor
when pin depth is adjusted and gaged and when the load is stored. Remove this
support bead from the sliding load before you connect the load for an electrical
calibration. Reinsert this support bead when you’ve finished using the sliding
load.
CAUTION
Circuitry inside the test set at the test ports may be destroyed if precautions are
not taken to avoid electrostatic discharge (ESD). During this procedure, the
center conductor of the sliding load is connected to the exposed center conductor
of the test port. Ground yourself to prevent electrostatic discharge.
CAUTION
The sliding load center conductor can be damaged if the sliding load is not in line
when mating the load to a connector. Always line-up the sliding load when
connecting or removing it from a connector.
1. Refer to
. Loosen the center conductor pull-back nut completely. Press the center
conductor cap to extend the center conductor of the sliding load beyond the end of the
connector.
2. Continue to press the center conductor cap and mate the center conductor of the sliding
load with the cable/test port connector’s center conductor.
3. Release pressure on the center conductor and mate the outer conductor of the sliding load
with the outer conductor of the cable/test port connector. Torque the connection with a 3/4
Содержание 85054B
Страница 1: ...Keysight Technologies 85054B 50 Type N Calibration Kit User s and Service Guide...
Страница 2: ......
Страница 3: ...85054B...
Страница 7: ...1 General Information...
Страница 15: ...2 Specifications...
Страница 20: ...2 6 85054B Specifications Electrical Specifications...
Страница 21: ...3 Use Maintenance and Care of the Devices...
Страница 29: ...85054B 3 9 Use Maintenance and Care of the Devices Gaging Connectors Figure 3 3 Gaging Male Type N Connectors...
Страница 43: ...4 Performance Verification...
Страница 46: ...4 4 85054B Performance Verification Recertification...
Страница 47: ...5 Troubleshooting...
Страница 52: ...5 6 85054B Troubleshooting Contacting Keysight...
Страница 53: ...6 Replaceable Parts...
Страница 56: ...6 4 85054B Replaceable Parts Introduction Figure 6 1 Replaceable Parts for the 85054B Calibration Kit...
Страница 57: ...85054B 6 5 Replaceable Parts Introduction Figure 6 2 More Replaceable Parts for the 85054B Calibration Kit...
Страница 58: ...6 6 85054B Replaceable Parts Introduction...
Страница 59: ...A Standard Definitions...
Страница 64: ...Index 85054B Index 4 part number 6 2 Z zeroing connector gage 3 8 3 10 3 12...