85054B
3-5
Use, Maintenance, and Care of the Devices
Cleaning Connectors
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you
reassemble or use it.
3.
Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to
. When cleaning a female connector, avoid snagging the swab on the center
conductor contact fingers by using short strokes.
c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you
reassemble or use it.
4.
Inspect
Inspect the connector again to make sure that no particles or residue are present.
Содержание 85054B
Страница 1: ...Keysight Technologies 85054B 50 Type N Calibration Kit User s and Service Guide...
Страница 2: ......
Страница 3: ...85054B...
Страница 7: ...1 General Information...
Страница 15: ...2 Specifications...
Страница 20: ...2 6 85054B Specifications Electrical Specifications...
Страница 21: ...3 Use Maintenance and Care of the Devices...
Страница 29: ...85054B 3 9 Use Maintenance and Care of the Devices Gaging Connectors Figure 3 3 Gaging Male Type N Connectors...
Страница 43: ...4 Performance Verification...
Страница 46: ...4 4 85054B Performance Verification Recertification...
Страница 47: ...5 Troubleshooting...
Страница 52: ...5 6 85054B Troubleshooting Contacting Keysight...
Страница 53: ...6 Replaceable Parts...
Страница 56: ...6 4 85054B Replaceable Parts Introduction Figure 6 1 Replaceable Parts for the 85054B Calibration Kit...
Страница 57: ...85054B 6 5 Replaceable Parts Introduction Figure 6 2 More Replaceable Parts for the 85054B Calibration Kit...
Страница 58: ...6 6 85054B Replaceable Parts Introduction...
Страница 59: ...A Standard Definitions...
Страница 64: ...Index 85054B Index 4 part number 6 2 Z zeroing connector gage 3 8 3 10 3 12...