3-10 85053B
Use, Maintenance, and Care of the Devices
Gaging Connectors
3-
For maximum accuracy, measure the connector a minimum of three
times and take an average of the readings. After each measurement,
rotate the gage a quarter-turn to reduce measurement variations
that result from the gage or the connector face not being exactly
perpendicular to the center axis.
e.
Compare the average reading with the observed pin depth limits in