85053B 3-9
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Procedures
Gaging 3.5 mm Connectors
1.
Select the proper gage for your connector. Refer to
for gage part numbers.
2.
Inspect and clean the gage, gage master, and device to be gaged. Refer to
and
earlier in this chapter.
3.
Zero the connector gage (refer to
a.
While holding the gage by the barrel, and without turning the gage
or the device, connect the gage to the gage master by
interconnecting the male and female connectors. Connect the nut
finger tight. Do not overtighten.
b.
Using an open-end wrench to keep the device body from rotating,
use the torque wrench included in the kit to tighten the connecting
nut to the specified torque. Refer to
for additional information.
c.
As you watch the gage pointer, gently tap the barrel of the gage to
settle the reading.
The gage pointer should line up exactly with the zero mark on the
gage. If not, adjust the zero set knob until the gage pointer lines up
exactly with the zero mark.
d.
Remove the gage master.
4.
Gage the device connector (refer to
):
a.
While holding the gage by the barrel, and without turning the gage
or the device, connect the gage to the device by interconnecting the
male and female connectors. Connect the nut finger-tight. Do not
overtighten.
b.
Using an open-end wrench to keep the device body from rotating,
use the torque wrench included in the kit to tighten the connecting
nut to the specified torque. Refer to
for additional information.
c.
Gently tap the barrel of the gage with your finger to settle the gage
reading.
d.
Read the gage indicator dial. Read
only
the black
±
signs;
not
the
red
±
signs.
Always hold a connector gage by the gage barrel, below the dial indicator.
This gives the best stability, and improves measurement accuracy.
(Cradling the gage in your hand or holding it by the dial applies stress to
the gage plunger mechanism through the dial indicator housing.)
Содержание 85053B
Страница 1: ...User s and Service Guide 3 5 mm Verification Kit Keysight 85053B ...
Страница 10: ...1 6 85053B General Information Preventive Maintenance 1 ...
Страница 18: ...3 2 85053B Use Maintenance and Care of the Devices Electrostatic Discharge 3 Figure 3 1 ESD Protection Setup ...
Страница 27: ...85053B 3 11 Use Maintenance and Care of the Devices Gaging Connectors Figure 3 2 Gaging 3 5 mm Connectors ...
Страница 38: ...3 22 85053B Use Maintenance and Care of the Devices Handling and Storage 3 ...
Страница 42: ...5 2 85053B Troubleshooting Troubleshooting Process 5 Figure 5 1 Troubleshooting Flowchart ...
Страница 50: ...6 6 85053B Replaceable Parts Replaceable Parts 6 Figure 6 1 Component Identification Sheet ...
Страница 56: ...85053 90028 Index Index 6 ...