Keysight DFB Laser Source Module User’s Guide
67
Specifications
5
Measurement
With wavelength meter. Averaging time given by wavelength meter,
≥
1 s.
Wavelength stability
The change of wavelength during given time span, expressed as ± half the
span between the maximum and the minimum of all measured
wavelengths.
Conditions
Uninterrupted DFB laser source module output power, constant
wavelength and power level settings, coherence control off, controlled
environment and time span as specified.
Measurement
With wavelength meter. Averaging time given by wavelength meter
≥
1 s.
Max. output power
The highest optical power obtainable at the end of a single-mode fiber
patchcord.
Conditions
At default wavelength.
Measurement
With power meter at the end of a single-mode fiber patchcord.
NOTE
The long-term wavelength repeatability can be obtained by taking the
wavelength repeatability and wavelength stability into account.