1
Performance Tests
84
Keysight 53210A/53220A/53230A Assembly Level Service Guide
Counter setup
1
Connect the equipment as shown in
Figure 1-11
Option 115/150 microwave channel 3 test setup
2
Press the
Preset
key to preset the counter.
3
If needed, perform an auto-calibration. (Refer to
4
Press the Channel
3
input key.
– Channel 3 is selected for test input.
O p t 0 1 0 U O C X O
L in e
1 0 0 - 2 4 0 V , 5 0 - 6 0 H z
1 0 0 - 1 2 7 V , 4 0 0 H z
9 0 V A M a x
U S B L A N
In t R e f O u t T r ig In
E x t R e f In G a te In /O u t
C h 1 O p t 2 0 1 C h 2 G P -IB
O p t 1 0 6 /1 1 5
IE
CS /
NMB-
001
N 1 0 1 4 9
IS M 1 -A
C U S
O p t 1 0 6
1 0 0 M H z - 6 G H z
+ 2 7 d B m M a x
!
O p t 3 0 0 B a t te r y
O p t 1 1 5
3 0 0 M H z - 1 5 G H z
+ 2 7 d B m M a x
!
FS725
Rubidium
Timebase
Ext Ref In
E8257D-520-UNW
Microwave
Signal
Generator
Ch 3