X, P, and K 11644A
5-5
Performance Verification
Performance Test
4. Set the averaging factor to 512.
5. At the adapter test port, perform a 2–port TRL calibration.
6. Turn on the calibration.
7. Connect the termination you wish to test to port 1.
8. Measure the return loss (S
11
) of the load.
9. After one complete measurement sweep, the displayed trace should look similar to that shown in
Figure 5-2
.
10. If necessary, update the trace.
11. Use a marker to determine the maximum value on the trace. This marker determines worst-case return
loss.
Figure 5-2 Typical Termination Return Loss
In Case of Failure
If a termination fails this test, clean all flanges and carefully reconnect the devices. Repeat the test. If the
termination fails again, replace it. Refer to
Содержание 11644A K
Страница 2: ......
Страница 3: ...X P and K 11644A i...
Страница 7: ...1 General Information...
Страница 12: ...1 6 X P and K 11644A General Information Preventive Maintenance...
Страница 13: ...2 Specifications...
Страница 19: ...3 User Information...
Страница 26: ...3 8 X P and K 11644A User Information Changing the Wavelength Shim Calibration Definition...
Страница 27: ...4 Use Maintenance and Care of the Devices...
Страница 36: ...4 10 X P and K 11644A Use Maintenance and Care of the Devices Handling and Storage...
Страница 37: ...5 Performance Verification...
Страница 52: ...5 16 X P and K 11644A Performance Verification System Operation Checks...
Страница 53: ...6 Troubleshooting...
Страница 57: ...7 Replaceable Parts...
Страница 58: ...7 2 X P and K 11644A Replaceable Parts...
Страница 61: ...X P and K 11644A 7 5 Replaceable Parts Introduction Figure 7 1 X Band Component Identification Sheet...
Страница 63: ...X P and K 11644A 7 7 Replaceable Parts Introduction Figure 7 2 P Band Component Identification Sheet...
Страница 65: ...X P and K 11644A 7 9 Replaceable Parts Introduction Figure 7 3 K Band Component Identification Sheet...
Страница 67: ...A Standard Definitions...