5.4 Calibration 

Corporate guidelines prescribe that all companies operating with a quality assurance system must regu-
larly calibrate their instruments and equipment. Calibration must ensure that the unit complies with the 
stipulated specifications. 

When should calibration be carried out? 

Calibration relates to the dosing volume of the pump as a function of the tube.  

Calibrate;  

  If a new tube type is used and added to the tube calibration store for later use. 

  If using culture media with greatly differing viscosities are used. 

  At regular intervals, e.g. every 3 months. 

  If the silicone tubing flexibility changes and results in a general volume drift. This is based on many 

autoclaving cycles, standard tube flexibility loss, age, UV exposure e.g.) 

NOTE 

The calibration procedure is described in Chapter "Calibration menu". 

5.5 Troubleshooting 

5.5.1 Overview 

The D

OSI

P

UMP 

1000 software incorporates troubleshooting routines which can be used to correct certain 

problems without action on the part of after-sales servicing technicians. 
If a problem occurs on the D

OSI

P

UMP 

1000, the unit attempts to correct the problem by providing the op-

erator with prompts on the graphical user interface. If these on-screen prompts are followed, the fault or 
error can then normally be remedied without action on the part of after-sales servicing technicians.  
Each error message with alarm function also contains a description of the fault or error. 

5.5.2 Fault 

clearance 

The table below can be used to help you to remedy certain problems which may occur during operation 
of the D

OSI

P

UMP 

1000. This table contains information on the symptom, its possible causes and the re-

medial action for the relevant fault or error involved. 

Errors which occur before operation 

Symptom 

Possible cause 

Corrective action 

Device does not start 

Plug not plugged in 

Master switch not switched on 

Master switch fuse defective 

No mains power 

Check power connection 

Switch on 

Check fuse.  
See details of description.  

Check power connection 

Replacing the device fuse 

Содержание 730'000

Страница 1: ...DosiPump 1000 User Manual Page 1 24 Version 1 01E DOSIPUMP DP 1000 DosiPump DP1000 Item 730 000 DosiPump DP1000 Fast Item 740 000 USER MANUAL 1 01 E Version 2010 2011...

Страница 2: ...ING 8 3 2 MAINS CONNECTION 8 3 3 INSERTING THE TUBE FITTING IN THE PUMP 8 3 4 METAL BAR FIXATION MAKING THE STURDY HOUSING A STABLE STAND 8 3 5 MAIN SWITCH 8 4 OPERATION OF THE DOSIPUMP 1000 9 4 1 FUN...

Страница 3: ...he Overview contains general information on the DOSIPUMP 1000 the intended application and the basic concept Safety Information The DOSIPUMP DP1000 safety information Commissioning The steps and prepa...

Страница 4: ...nge in only 1 unit Precision and reproducibility are depending the ratio between tube and desired volume The smaller the tube the higher the precision and reproduci bility but the smaller the pump spe...

Страница 5: ...epts tube sizes from 2x6mm for small doses up to a double tube set of 8x12mm for fast dosing DosiPump is build in a stable stainless steel housing offering IP64 splash water resistance DosiPump DP1000...

Страница 6: ...chemicals DosiPump is not an EX Approved WARNING We are unable to guarantee operability and functional safety of the unit if the ambient conditions applicable to the location are not complied with 2 2...

Страница 7: ...her purposes 2 Maintenance work and repair work may be carried out only by qualified staff 3 The unit must be installed in such a manner that the On Off switch is easily accessible at all times 4 Shou...

Страница 8: ...ese values Connect the power cable of the DOSIPUMP 1000 only to a fully earthed mains power supply system in order to avoid electrical shock Do not use adapters for the plugs and do not remove the ear...

Страница 9: ...Pump speed ml min the pump direction Antidrop angle defined pump head pullback pause before next dose 4 4 All Programmes require a calibrated tube DosiPump allows 8 Tube sets to be cal ibrated and sel...

Страница 10: ...to continue 2 3 4 DosiPump allows direct dosing straight from the main screen continuous pump mode No need for previous program setting The speed and the pump direction can be selected Select Pump di...

Страница 11: ...irection front backside Anti Drop Angle in Delay between singe doses sec Start Program 4 5 Select the number of repetitive dosings or endless repeats manual stop required Tube Filling option before pr...

Страница 12: ...fore DosiPump starts the selected program the cycle repetition screen appears automatically Select number or cycles repetition of complete program Alternatively select Endless repetitions repeats unti...

Страница 13: ...opens Selects the tube 1 8 Each single Dosing Step consists of all individual variables a dosing can consist of Opens the Edit Mode of the selected step Start Stop dispensing this program Volume Pump...

Страница 14: ...se as possible This based on the physical difference any tube offers regarding pump expansion relaxation between the pump rolls Note The pump direction in the calibration mode is CCW Info The pump cal...

Страница 15: ...e Touch Screen does not react fast and appropriate anymore it may be necessary to recalibrate the touch screen Touch at the top left hand corner Follow the at the bottom right hand corner 4 7 4 Displa...

Страница 16: ...three parts water Deionised water Soft lint free cleaning cloth and paper tissues WARNING Do not use cleaning agents containing acetone or other aggressive cleaning agents nitrocellulose thinners WAR...

Страница 17: ...porates troubleshooting routines which can be used to correct certain problems without action on the part of after sales servicing technicians If a problem occurs on the DOSIPUMP 1000 the unit attempt...

Страница 18: ...urring during operation Symptom Possible cause Correction Fault signal Pump blocked The culture medium in the tube is too viscous You have used a tube which is too large Remove the tube and run the pu...

Страница 19: ...act 1 Pin 3 Footswitch contact 2 Pin 2 29 V The two footswitch contacts are designed as normally open contacts NO 2 PC printer connection pin assignment RS232 Version P PI Pin 1 RxD Pin 2 TxD Pin 3 GN...

Страница 20: ...everse angle of pump head Display 240 x 128 pixels with LED backlighting Sound pressure level 79 dB A Pollution severity II Safety class IP 64 Ambient temperature 5 C to 45 C also during transport Uni...

Страница 21: ...0 Stage bar stainless steel 101052 Silicone tubing 3x7 mm roll of 25m 101054 Silicone tubing 5x9 mm roll of 25m 101055 Silicone tubing 6x10 mm roll of 25m 101056 Silicone tubing 7x10 mm roll of 25m 10...

Страница 22: ...4 Version 1 01E 6 2 Reproducibility and Precision Data from BioTool test lab June 25th 2010 DP1000 Serial 730 13 00014 DP1000 Fast Serial 740 13 00016 Test with tab water No evaporation protection oil...

Страница 23: ...automatic updates and product infor mation from BioTool AG 6 5 Contacts Feedback Sales Marketing Manufacturing site Please place your feedback with you local supplier your local service partner BioTo...

Страница 24: ...DosiPump 1000 User Manual Page 24 24 Version 1 01E...

Отзывы: