23
5.3 Airflow Sensor installation.
INSTALLATION NOTES :-
The back connection spigot of the sensor is designed to accept a
1”
OD tube which should be connected to
the inner chamber of the fume hood. (This tube and fittings is known as the " vent kit " )
The ideal position for the end of the 1
”
tube for most fume hoods is
4”
back from the sash glass
and
4”
higher than the normal sash opening height through the inner side
Adapter through
Inner side wall
Front of
side post
AFA500
Mounted on
front of
side post
AA300
Mounted on front
of side post
Front of
side post
Adaptor through
Inner side wall