13 
 

3.1 Alarm Configuration / Calibration 
 

 

 

The alarm is supplied with a factory configuration. 
The only part of the configuration that can be changed is the setting for the Visor Open repeat timer time delay -- 
if required this can be factory changed prior to ordering the monitor. 
 
The alarm has two modes for the calibration of the airflow alarm point. These two modes are selectable via a  

‘jumper connection’ on the back of the alarm PCB.

 

 
The jumper selection calibration modes are :- 
 

a. With Jumper J2 connected -   The airflow is reduced to the alarm point mechanically and this airflow is 

‘captured’ by the alarm. The airflow is then restored to the normal operating value and the Low Air alarm will 

activate if the airflow subsequently falls to the alarm point 

 

b. With JumperJ2 not connected -  This is a two point calibration method. The airflow is set to the normal 

operating value and this value is ‘captured’ by the alarm and taken to be 100%. The airflow is then increased 
to one and a half time the normal operating value and this value is ‘captured’ by the alarm and taken to be 

150%. The Low Air alarm will then activate if the airflow subsequently falls to 80% of  the normal operating 
value. 
   

 
 

See  ‘Quick Start Installation’ below for details of the Calibration procedures

 

 
 
 
 
 

3.2 Start Up 

 

When unit is powered up  the following sequence of events occur :- 
 
 
1.

 

 The 12V DC power is applied to the airflow sensor and the alarm then performs a self-test on the functions , 
LEDs and audible alarm ( approx 2 sec ) and then initiates a delay timer of 30 secs to allow the airflow sensor 
to stabilize. 

 
2.

 

During the whole of the 30 sec period all alarms and relay outputs are inhibited and the Red & Green LEDs 
will be permanently ON. 

 
3.

 

At the end of the 30 sec delay the unit performs one of two options :- 

 

 

a. If the alarm calibration has been previously completed 

 the unit goes to   

     normal operating mode ( Run ) 
 
b. If the unit has not been calibrated -- the Red & Green LEDs will flash on/off and 

   

 the audible alarm will be muted. It is then possible to press the Enter button for  

   5 secs and go into the calibration mode 

– (See  ‘Quick Start Installation’ below for 

          details of the Calibration procedures

 

 
 
 
 
 
 

Содержание Air Alert 300

Страница 1: ...1 AIR ALERT 300 FUME HOOD VELOCITY ALARM Installation and Operating Manual Issue 01 February 2019 ...

Страница 2: ...or Display Panel 9 2 2 General Description 10 2 3 External Connections 11 3 Functions and Operation 12 3 1 Configuration and Calibration 13 3 2 Start Up 13 3 3 Events and Actions 14 3 4 Quick Start Installation 15 3 5 Calibration Notes 17 4 Components 18 4 1 Airflow Monitor 19 5 Installation 20 5 1 Installation of Standard Components 21 5 2 Airflow Monitor Installation 22 5 3 Airflow Sensor Instal...

Страница 3: ...3 1 SAFETY ...

Страница 4: ...ng up the equipment or performing analysis or maintenance procedures strictly follow the instructions provided 1 2 Warning Notices Within this User Guide WARNINGS are used to highlight information or instructions that must be followed in order to avoid personal injury to yourself or other people in the vicinity eg switch off the mains voltage before any maintenance WARNINGS appear as below WARNING...

Страница 5: ...must be done by a skilled person who is aware of the hazards involved Whenever circumstances arise that mean an Airflow Monitor may be unsafe make it inoperative In particular an Airflow Monitor may be unsafe if it Shows visible damage Fails to perform correctly Has been subjected to severe transport stresses Has been subjected to prolonged storage in unfavorable conditions 1 5 Environmental Condi...

Страница 6: ...ctive conductor terminal is likely to make the equipment dangerous Intentional interruption of the protective conductor is prohibited WARNING If the mains power supply has to be replaced ensure that the replacement power supply is appropriately rated and approved for the intended use WARNING To prevent potential personal injury and damage to the equipment switch OFF all components in the system an...

Страница 7: ...ion occurs Occasionally however temporary conductivity caused by condensation must be expected 1 8 EMC Compliance EC Directive The Airflow Monitor System is designed and tested to meet the requirements of the EC directive 89 336 EEC and 93 68 EEC and complies with the EMC standard EN61326 EMC standard for electrical equipment for measurement control and laboratory use and EN55011 ISM Class A RF em...

Страница 8: ...8 2 Overview of the Airflow Monitor ...

Страница 9: ...9 2 1 Operator Display Panel Enter and Set buttons for Calibration ENTER also used as Mute button for audible alarm LED Indicators ...

Страница 10: ...tion Night Set back when the Night Setback input is activated the Audible alarm will be muted and the Green LED will flash on off ENTER the alarm has an Enter button this is multi functional as follows Press Enter momentarily when Low Air alarm is sounding will mute the alarm Press Enter momentarily when Visor Open alarm is sounding will mute the alarm and initiate a repeat timer that will re soun...

Страница 11: ...n activation This input is configured as LOW AIR ALARM Visor Open Input a Connection point for Visor micro switch Switch contact to close and remain closed in Visor Open condition b Connection point for Visor proximity switch Switch contact to close and remain closed in Visor Open condition Note Use input a OR input b for the Visor alarm Power supply low voltage DC power supply 15V DC ...

Страница 12: ...12 3 Functions Operation ...

Страница 13: ... half time the normal operating value and this value is captured by the alarm and taken to be 150 The Low Air alarm will then activate if the airflow subsequently falls to 80 of the normal operating value See Quick Start Installation below for details of the Calibration procedures 3 2 Start Up When unit is powered up the following sequence of events occur 1 The 12V DC power is applied to the airfl...

Страница 14: ... 1 sec Audible can be muted via Enter pushbutton this silences the alarm and initiates a repeat timer factory set to 5 mins After the delay time the alarm re sounds and can be re muted During this time the Red LED flashes on off Reset when Visor closed to safe position and input de activated Night set back When input configured as Night set back is activated Green LED on Flashing Audible alarm mut...

Страница 15: ... the alarm point level b Press and hold the Enter button for 5 secs to go into the calibration mode This is indicated by both Red and Green LEDs flashing on off together with the audible alarm sounding Beep on off 4 times every 1 sec c To initiate the alarm point calibration press and hold the ENTER and the SET buttons at the same time The unit will then sample the airflow for a 5 sec period durin...

Страница 16: ... SET buttons to repeat the airflow sampling d The higher airflow calibration mode is indicated by both Red and Green LEDS flashing on off together with the audible alarm sounding Beep on off for 2 secs every 10 secs Adjust the fan speed until the face velocity rises to one and a half times the normal operating value used in a above using a calibrated instrument to check the value To initiate the h...

Страница 17: ...nt calibration is ideal It is recommended that the fan is set to give an alarm velocity of 0 3 m sec 60 fpm minimum and 0 5 m sec 100 fpm maximum This considers a normal operating velocity of approximately 0 75 m sec 150 fpm Using the two point capture method is ideal for low air alarm calibration where the alarm point is required to indicate a low operating velocity typically 80 of the normal ope...

Страница 18: ...18 4 Components ...

Страница 19: ...19 4 1 Airflow Monitor components 1 Air Alert AA300 Airflow Monitor 1 Plug in type low voltage power supply with 5M Cable TEL ...

Страница 20: ...20 5 Installation ...

Страница 21: ...s are not possible though the principles outlined below should remain valid in all cases The airflow monitor can be mounted on either side of the Fume Hood 1 Fit the AA300 monitor to the Fume Hood using the cut out details provided with the unit Ensure that the monitor fits into the cut out before securing with the fixing screws 2 Fit the airflow sensor hose to the Fume Hood using the cut out and ...

Страница 22: ...22 5 2 Airflow Monitor Installation Alarm Panel Dimensions Fume Hood Side Airfoil Front View Alarm Panel Cut Out ...

Страница 23: ...od This tube and fittings is known as the vent kit The ideal position for the end of the 1 tube for most fume hoods is 4 back from the sash glass and 4 higher than the normal sash opening height through the inner side Adapter through Inner side wall Front of side post AFA500 Mounted on front of side post AA300 Mounted on front of side post Front of side post Adaptor through Inner side wall ...

Страница 24: ...24 5 4 Typical connection diagram TEL ...

Страница 25: ...25 6 Warranty ...

Страница 26: ...epaid and will be returned at Seller s risk of loss Buyer s factory prepaid The foregoing is IN LIEU OF all other warranties and is subject to the conditions and LIMITATIONS stated herein NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE THE EXCLUSIVE REMEDY OF THE USER OR PURCHASER AND THE LIMIT OF LIABILITY OF SELLER FOR ANY AND ALL LOSSES INJURIES...

Отзывы: