background image

Coating Thickness Tester

L-500

Operating Manual

Thank you for purchasing this product.

Please read the operating manual carefully and use this product properly.

Содержание L-500

Страница 1: ...Coating Thickness Tester L 500 Operating Manual Thank you for purchasing this product Please read the operating manual carefully and use this product properly...

Страница 2: ...If overvoltage is applied the unit will be heated and it may cause malfunction fire or electric shock Make sure to ground the power cable With the 2P outlet use a conversion plug to ground In the case...

Страница 3: ...escription of display and operation keys 8 4 Measurement Preparation 9 5 Measurement Mode 11 6 About Statistics Function 12 7 Various settings 14 8 Battery printer paper replacement 39 9 Data output 4...

Страница 4: ...n magnetic metals If an electromagnetic coil through which a constant low frequency current flows is brought into the proximity of iron magnetic metal the number of magnetic flux lines through the coi...

Страница 5: ...ate for Fe probe aluminum substrate for NFe probe calibration foil x 6 battery 1 5V AA alkali 8 AC adapter power cable POC00 probe adapter strap printer paper 2 surface protective sheet 3 calibration...

Страница 6: ...attery compartment for main unit Data output terminal USB Mini B AC adapter socket Open close button Printer paper cover Printer paper cutter Display Probe holder Operation keys Probe connector Batter...

Страница 7: ...adapter Strap Printer paper 2 Protective sheets 3 At the time of shipment one roll is set in the main unit Calibration foil case Carrying case Easy Guide for Calibration Operating Manual Electromagne...

Страница 8: ...DELETE key Erases the measured data from memory This key is also used to erase the entered values and characters at entry of them POWER key Turns on off the power of the main unit To turn off the pow...

Страница 9: ...open it Install four 1 5 V AA alkaline batteries with attention to the plus and minus directions Be careful not to damage the lug of the battery case lid when opening or closing the lid When this test...

Страница 10: ...t part of the probe and make an insertion The probe is properly secured with a sound of click Secure the probe cable to the cable holder The probe grip can be secured to the probe holder Probe detachi...

Страница 11: ...rement should be performed after the tip of the probe is separated from the measurement surface by 10 mm or more To measure a pipe shaped object or a flat surface continuously a probe adapter can be u...

Страница 12: ...sured data classification which is a data classification that stores arbitrary blocks together Non consecutive blocks can be grouped together because a group with which each block can be registered ca...

Страница 13: ...ter measurement and the data that is not suitable for statistical calculation can be deleted from the list This deletion can be used when obviously abnormal data is deleted by viewing the list afterwa...

Страница 14: ...lower limits P 22 5 Unit P 23 6 Measurement mode P 24 Statistics menu 7 Statistics P 25 8 New block P 27 9 Group function P 28 0 Delete data P 30 Data memory P 31 Data output P 32 Setting menu Auto p...

Страница 15: ...lect the measurement menu icon with the and keys and then press the key S u b s t r a t e c a l F o i l c a l i b r a t i o n Application L i m i t s U n i t Measurement mode BACK 3 Select Substrate c...

Страница 16: ...e is up to 1 200 m Use the attached substrate to easily check the accuracy of the instrument Values of the provided calibration foils are not the exactly same as the values described in the table but...

Страница 17: ...SET 2 Perform substrate calibration P 15 3 Press the SET key on the measurement screen P 14 2021 01 15 1 4 2 6 N 05535 APP 01 TEST 1 GRP 01 B 001 Bn 05535 Av g 1 2 3 Fe BLOCK GROUP APP SET Press and...

Страница 18: ...e substrate and measure about 5 times Measure in order from the thinnest calibration foil to be used If you make a measurement error you can delete the previous value by pressing the DELETE key W h e...

Страница 19: ...screen of step 5 9 Measure the calibration foil 3rd foil Place the calibration foil 500 m on the substrate and measure about 5 times Perform operations in the same manner as step 7 Cal APP 01 TEST 1...

Страница 20: ...at the time of shipment The current application number is displayed as APP 01 The set application number will be kept intact even after the power is turned off until it is changed When selecting Chan...

Страница 21: ...1 TEST 1 APP 02 APP 03 APP 04 APP 05 APP 06 APP 07 APP 08 BACK ENTER By pressing the DELETE key during selection of application number you can delete the saved name calibration result and the upper an...

Страница 22: ...ed here but operation of the lower limit can be done in the same manner Select Enable or Disable with the and keys and then press the ENTER key L i m i t s APP 01 U p p e r l i m i t 0 0 um Enable Dis...

Страница 23: ...nd then press the ENTER key S u b s t r a t e c a l F o i l c a l i b r a t i o n Application L i m i t s U n i t Measurement mode ENTER 4 Select the unit to display Select the unit to be displayed wi...

Страница 24: ...ions 1 Press the SET key on the measurement screen P 14 2 Select the icon Select the measurement menu icon with the and keys and then press the key S u b s t r a t e c a l F o i l c a l i b r a t i o...

Страница 25: ...in each block can be displayed on the screen and printing of them is available You can also display the results by pressing the BLOCK RESULT key Data statistics Statistical calculation results for dat...

Страница 26: ...GRP 06 Input number 1 50 BACK ENTER The current group number will be displayed on the group function screen If you make an input error you can delete the number by pressing the DELETE key B l o c k s...

Страница 27: ...n press the key S t a t i s t i c s New block Group function Delete data Data memory Data output BACK 3 Select Block update Select Block update using the and keys and then press the Enter key S t a t...

Страница 28: ...t necessary to change it press the BACK key Group GRP 01 Change number Change name App list BACK ENTER T h e c u r r e n t g r o u p n u m b e r i s displayed as GRP 00 When selecting Change number Af...

Страница 29: ...desired group number with the and keys and then press the ENTER key GRP 01 TEST 1 GRP 02 GRP 03 GRP 04 GRP 05 GRP 06 GRP 07 GRP 08 BACK ENTER 5 The screen will return to the screen of step 3 If any of...

Страница 30: ...4 Select the item to delete Select the item to delete with the and keys and then press the ENTER key Delete data A l l d a t a Select data BACK ENTER When selecting Delete all data A c o n f i r m a t...

Страница 31: ...item to set with the and keys and then press the ENTER key If it is not necessary to change it press the BACK key Data memory Enable Disable BACK ENTER At the time of shipment Enable is set When Enab...

Страница 32: ...press the key S t a t i s t i c s New block Group function Delete data Data memory Data output BACK 3 Select Data output Select Data output with the and keys and then press the ENTER key S t a t i s t...

Страница 33: ...uto power off with the and keys and then press the ENTER key Auto power off Backlight P r i n t d e n s i t y Time and date D i s p l a y s e t t i n g Software version Service menu ENTER 4 Select whe...

Страница 34: ...y Time and date D i s p l a y s e t t i n g Software version Service menu BACK 3 Select Backlight Select Backlight with the and keys and then press the ENTER key Auto power off Backlight P r i n t d...

Страница 35: ...P r i n t d e n s i t y Time and date D i s p l a y s e t t i n g Software version Service menu BACK 3 Select Print density Select Print density with the and keys and then press the ENTER key Auto pow...

Страница 36: ...d time with the and keys and then press the ENTER key Auto power off Backlight P r i n t d e n s i t y Time and date D i s p l a y s e t t i n g Software version Service menu ENTER 4 Select the settin...

Страница 37: ...ENTER key Auto power off Backlight P r i n t d e n s i t y Time and date D i s p l a y s e t t i n g Software version Service menu ENTER 4 Select a display method on the measurement screen Select an...

Страница 38: ...and date D i s p l a y s e t t i n g Software version Service menu BACK 3 Select Software version Select Software version with the and keys and then press the ENTER key Auto power off Backlight P r i...

Страница 39: ...ll not occur Close the printer paper cover with the tip of the printer paper sticking out as shown in the photo If the following icon is displayed check by following the respective procedures Name Ico...

Страница 40: ...0A Example Measurement result when reached the limit 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 Output data Character N 0 0 0 0 6 1 2 3 u m N 00006 12 3 um ASCII code...

Страница 41: ...er S e t U L I M 1 7 3 9 u m Set U LIM 1739 um ASCII code 20 20 20 53 65 74 20 55 2E 4C 49 4D 20 20 20 31 37 33 39 20 20 20 20 75 6D 20 20 0A Character S e t L L I M 2 5 8 u m Set L LIM 25 8 um ASCII...

Страница 42: ...libration foils are precisely measured in thickness If you make calibrations using damaged or bent foils you will not be able to make accurate measurements Pay particular attention to the wear of the...

Страница 43: ...n is made according to the measurement target and the thickness you want to measure P 16 Foil calibration Cannot print Power supply printer Is the power of the printer on Press the PRINTER POWER key W...

Страница 44: ...In this case contact your dealer or us Under Range Error 104 Exceeding the lower limit of measurement range Check the measurement range of the probe in use Refer to the instruction manual provided wit...

Страница 45: ...45...

Страница 46: ......

Страница 47: ......

Страница 48: ...ce The appearance and or representations of the products and parts depicted in this user manual may not appear exactly as their actual counterparts but this does not affect their operation or function...

Отзывы: