![Kett L-500 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/kett/l-500/l-500_operating-manual_1958599024.webp)
24
6
Measurement Mode
Either of "Hold mode" that fixedly displays a measurement value and "Continuous
mode" that displays measurement value by continuous monitoring can be set.
* The measurement mode at the time of calibration is fixed to "Hold mode".
* In "Continuous mode", press the tip of the probe against the measurement point and
press the [HOLD] key when the measurement value is stabilized. The displayed value at
this time will be stored as measurement data and will be printed when the power of the
printer is on.
This is useful when measuring a complex shape where measurement values tend to be
unstable.
Operating procedure
Display
Precautions
1. Pr e s s t h e [ S E T ] key o n t h e
measurement screen (→P.14).
2. Select the icon [ ].
Select the measurement menu icon
[ ] with the [▲] and [▼] keys, and
then press the [►] key.
▶
S u b s t r a t e c a l .
▼
F o i l c a l i b r a t i o n
A p p l i c a t i o n
L i m i t s
U n i t
Measurement mode
BACK
▲
▼
▶
3. Select “Measurement mode”.
Select "Measurement mode" with
the [▲] and [▼] keys, and then
press the [ENTER] key.
◀
S u b s t r a t e c a l .
F o i l c a l i b r a t i o n
A p p l i c a t i o n
L i m i t s
U n i t
>> Measurement mode
◀
▲
▼
ENTER
4. Select an item to set.
Select a measurement mode to be
set with the [▲] and [▼] keys, and
then press the [ENTER] key.
If it is not necessary to change it,
press the [BACK] key.
◀
○
Measurement mode
>> Hold mode
Continuous
mode
BACK
▲
▼
ENTER
* At the time of shipment, it is set to
"Hold mode".
* When “Hold mode” is set, the icon (
)
will appear on the display.
5. The screen will return to the
screen of step 3.