18

9. Zero-adjustment

●  What is zero-adjustment?

If errors occur in the measurement results due to light source consumption or other 

causes, read the absorbance obtained from reflection on the zero-adjustment plate. 

The measurement result will be stabilized by setting the absorbance as zero.

Make sure to perform this operation when making calibration curve or in the regular 

measurement as well.

Also, the zero-adjustment is required after the power is turned on.

If performing zero-adjustment without cycling the power, operate the following:

 Attach the zero-adjustment plate.

  Open the light shielding cover, and set the zero-adjustment plate onto the rotation table of the 

unit. Then, close the cover.

 

  

 Perform zero-adjustment.

  Press the [ZERO] key. The zero-adjustment screen will be displayed.

 

Z E R O

X 1

0 . 0 0 0 1

X 2

− 0 . 0 0 0 1

X 3

0 . 0 0 0 2

X 4

0 . 0 0 0 3

X 5

− 0 . 0 0 1 2

X 6

0 . 0 0 0 8

  Press the [ENTER] key. The zero-adjustment will be started.

 

  When the zero-adjustment is completed, the initial screen will be restored automatically.

  In the regular measurement mode, the previous measured value will be displayed.

 

0 1

F L O U R

1

M O I S T U R E

2

P R O T E I N

1 3. 3

8. 8

3

C O M P 3

4

C O M P 4

1. 4

7 6. 1

 

*   The zero-adjustment is to calibrate the absorbance obtained by measuring the zero-adjustment plate to 

zero. It does not make the composition measurement value zero.

Содержание KB-270

Страница 1: ...NIR Composition Analyzer KB 270 Operating Manual Thank you for purchasing this product Please read the operating manual carefully and use this product properly...

Страница 2: ......

Страница 3: ...3 7 Making the calibration curve 14 7 1 Setting the measurement range and the number of samples 14 7 2 Setting the sampling times 14 7 3 Preparing the samples 14 7 4 Moisture preparation 15 7 5 Measur...

Страница 4: ...dor Do not allow the unit to come in contact with water This unit is not waterproof Do not allow water or other liquids to get into the unit s enclosure as this may lead to electric shock or malfuncti...

Страница 5: ...t sunlight electromagnetic interference corrosive gases or large amounts of dust Place the unit on a flat and stable surface where it will not be subjected to significant vibration during use When mov...

Страница 6: ...composition measurements It measures the near infrared absorption of six wavelengths of the measuring object and converts the obtained data to the composition The multiple compositions four compositi...

Страница 7: ...ual this document PC software USB cable VZC61 Using the supplied PC software various operations such as zero adjustment measurement setting the calibration curve device setting and maintenance can be...

Страница 8: ...content Channel number Channel name Composition Absorbance Operating temperature range 5 to 35 C no condensation Power supply 100 120 V AC 220 240 V AC 50 60 Hz Power Consumption 50W Dimensions and w...

Страница 9: ...A key Used for measurement TABLE key Used to set ON OFF of the rotation table ZERO key Used for zero adjustment EXIT key Used to exit the operation MODE key Used to select the measurement mode CH key...

Страница 10: ...justment will proceed while displaying the absorbance in real time When the calibration is completed the initial screen will be restored If the power is cycled repeat the operation from step 6 Make su...

Страница 11: ...surement error tends to occur depending on the light absorption characteristics of the plastic bag 3 Polyethylene sheet only for moisture Rice ball boiled noodle and uncooked noodles etc For large sol...

Страница 12: ...n When rotating the table using a 90 mm of Petri dish use the sample cell holder Also use the light shielding cover to prevent falling off When using the polyethylene sheet prevent the sheet from bein...

Страница 13: ...e saturated sampling amount If the sampling amount is sufficiently prepared set the sampling amount more than the absorbance saturated sampling amount Refer to P 30 Remarks if the absorbance saturated...

Страница 14: ...This is called calibration curve The calibration curve differs depending on the measuring object Therefore it is required to make calibration curve per object in advance This unit is capable of makin...

Страница 15: ...the obtained data as the standard composition As the over drying method the official method or a measurement using an infrared moisture analyzer is available Note that the sample cannot be reused in...

Страница 16: ...ing 1234 Present ON 1 2 3 4 Present OFF Absent Setting not available When changing the measurement channel the channel that has no coefficient or the one whose display setting of all four compositions...

Страница 17: ...7 6 X 3 0 5 1 5 4 X 4 0 4 1 8 1 X 5 0 1 9 8 4 X 6 0 1 2 9 6 Set the sample to cover up the measurement window The continuously measured absorbance will be displayed Set the smoothing value as require...

Страница 18: ...he light shielding cover and set the zero adjustment plate onto the rotation table of the unit Then close the cover Perform zero adjustment Press the ZERO key The zero adjustment screen will be displa...

Страница 19: ...essing the ENTER key will enable the setting Set the item referring to the corresponding pages Display Contents Page 1 C H P A R A M E T E R Setting the channel parameter P 20 2 C H D U P L I C A T E...

Страница 20: ...A S N T 2 C O M P P A R A M E T E R C A N C E L Enter the smoothing value of continuous measurement Select the smoothing value 0 to 5 using key and press the ENTER key to set A channel identification...

Страница 21: ...set To enter or edit Move the cursor using the EXIT and ENTER keys Input the composition name from the 1st character using the numeric keypad and move the cursor one by one with the Enter key Up to 8...

Страница 22: ...N C E L Enter the offset value To correct enter the value in the same procedure as step 11 Enter 0 if correction is not needed 1 0 1 F L O U R C O M P P A R A M E T E R 1 M O I S T U R E O F F S E T...

Страница 23: ...uplication destination channel Move the cursor to the destination of channel duplication with key and press the ENTER key to set Press the ENTER key when the confirmation screen is displayed The menu...

Страница 24: ...0 2 F L O U R 2 12 3 4 0 3 C O R N 1 12 4 0 4 C O R N 2 1 0 5 B U T T E R 1 2 3 0 6 F L O U R 1234 0 7 Select the channel to initialize Move the cursor to the channel to initialize with key Then press...

Страница 25: ...M is selected Select USB or LAN using key and press the ENTER key to set 4 C O M S E T T I N G C O M U S B X P O R T O K C A N C E L 4 C O M S E T T I N G M A C A D D R E S S I P A D D R E S S O K Che...

Страница 26: ...5 P A S S C O D E N E W C O D E 1 2 3 4 Set the pass code After four digits are input the cursor will move to the right end automatically Pressing the ENTER key will restore the menu screen after savi...

Страница 27: ...d off Be careful of the edges inside the main unit Otherwise it may cause an injury etc 11 Maintenance Step Operation Rear of the main unit Turn off the power and disconnect the power cable The lamp h...

Страница 28: ...11 3 Consumable parts To use the unit in the best accuracy it is recommended to replace consumable parts at an interval of two years Other than the light source lamp do not replace the consumable part...

Страница 29: ...k the ambient temperature E 05 Sample low reflection The reflectivity of sample is low and the signal level is below the requirement When sample is not set on the measurement window Set the sample and...

Страница 30: ...absorbance saturated sampling amount While the near infrared from the bottom passes through the sample the displayed value will continue to fluctuate If the following is met making calibration curve...

Страница 31: ...31...

Страница 32: ...rts depicted in this user manual may not appear exactly as their actual counterparts but this does not affect their operation or functionality This user manual was intended to be written as clearly an...

Отзывы: