12. Error display
The following errors are displayed when there is a problem with the unit or the
use condition. When an error is displayed, check the status and take measures
following the procedures described below.
Error display
Description
Remedy
E-01
System parameter memory error
For purchasing the spare parts, please contact the
vendor.
E-02
Arithmetic error
E-03
Motor rotation error
E-04
Sensor internal temperature error Check the ambient temperature.
E-05
Sample low reflection
(The reflectivity of sample is low,
and the signal level is below the
requirement.)
● When sample is not set on the measurement window
S e t t h e s a m p l e, a n d m a ke s u r e t h a t t h e
measurement value is displayed.
● When zero-adjustment plate is set
The light source lamp may have blown. For
purchasing the spare parts, please contact the
vendor.
E-06
Channel parameter not set
Set an appropriate channel parameter.
E-07
Sensor data input error
For purchasing the spare parts, please contact the
vendor.
E-08
Channel parameter memory error
E-10
Device data memory error
E-11
Rotation table rotation error