The following error are displayed when this instrument or usage conditions are not normal. In such cases, please
contact us because repair may be necessary. This is also true should other error messages be displayed.
8. Error Display
Error Display
Description
Remedy
001
Problem with the temperature measurement circuit.
Contact to our distributor.
002
Problem with the capacitance measurement circuit.
003
Problem with the weight measurement circuit.
0 1 1
The measurement of empty capacitance is too high.
0 1 3
Main unit temperature is too high: over 60 °C.
Use the tester in the operating temperature
of 0 to 40 °C.
014
Main unit temperature is too low: under -10 °C.
015
Main unit board temperature is too high: over 60 °C.
016
Main unit board temperature is too low: under -10 °C.
0 1 7
Tare weighing is unstable for more than 20 seconds.
Use the tester in conditions that wind is
not strong and no vibration is sensed.
018
Tare weight is too high.
Contact to our distributor.
019
Tare weight is too low.
021
Sample weighing is unstable for more than 20 seconds.
Use the tester in conditions that wind is
not strong and no vibration is sensed.
024
Sample temperature is too high: over 60 °C.
Try measurement after the samples
a re the r ma lly homoge nize d in the
measurement environment.
025
Sample temperature is too low: under -10 °C.