12
(3) After the decimal point blinks for approx. 5 seconds, the
measuring count and moisture content are displayed.
Display example : 01 PKSD 1TIMES 13.5%
(Product number : 01 PKS (Dura),
1st measurement, moisture content 13.5%)
The measuring count is displayed from 1 to 9.
The 10th measuring count is displayed as 1.
[Note 1] For samples with high moisture content, the moisture
content dif ference among individual grains is
large. While the moisture content can be displayed,
measurement precision is not as good as lower
moisture content samples.
[Note 2] When the moisture content is below the measurement
range, "Lo" is displayed, and when the moisture
content is above the range, "Hi" is displayed.
[Note 3] When the sample is loaded into the measuring unit,
some sample may remain on the electrode at the
center of the measuring unit. In such cases, use the
attached brush to put the sample into the measuring
unit within 5 seconds of the decimal point blinking. If a
measurement is performed while the sample remains
on the electrode, an error may occur.
(example of below measurement range)
(example of above measurement range)