Display
Explanation
1. By using
keys, select
the “Adjustment” setting.
2. Press the
key.
A n d s e l e c t i t e m b y u s i n g
key.
3. Press the
key.
And change bias value by using
keys.
4. Press the
key.
While you press the
key, you can
increase and decrease the values
quickly with
keys.
If you want to set another choices,
repeat operation procedure 2~4.
If you press the
R T N
key, the initial
screen will be displayed.
Note : Sometimes, BIAS adjustment doesn’t run correctly - For example when the windows on sample
case is dirty or when the sample becomes rather damaged. If it get dirty, measure after wiping
the dirt off of it. In case that the sample is damaged, you would be replaced new standard sample.
Note : BIAS value is also acceptable with minus, instead of plus.
BIAS
= AV. Standard - AV. Measurements
- 0.5
=
6.1
-
6.6
In case of BIAS value is mius -0.5
[Adjustment]Application 1 Wheat
BIAS
/ Protein
− 0 0 . 1 %
/ Moisture
+ 0 0 . 1 %
BIAS Auto-adjustment with STD. sample.
Select the item.
[Adjustment]Application 1 Wheat
BIAS
/ Protein
− 0 0 . 2 %
/ Moisture
+ 0 0 . 1 %
BIAS Auto-adjustment with STD. sample.
Change the setting.
[Adjustment]Application 1 Wheat
BIAS
/ Protein
− 0 0 . 2 %
/ Moisture
+ 0 0 . 1 %
BIAS Auto-adjustment with STD. sample.
Select the item.
[ Measurement
]
[ Application
] 1 Wheat
[ Display
]
[ Adjustment
]
Serial number
[ Time
]
AB01234
[ Communication ]
VER 1 .08
2 0 0 6 − 0 8 − 3 1
1 3 : 5 9