Instructions for use elements™ 8:1
5 Operation | 5.1 Attaching the medical device
16 / 32
5 Operation
5.1 Attaching the medical device
CAUTION
Detachment of the medical device during treatment.
Injury or property damage
A medical device that is not properly locked in place can become disconnected
from the motor coupling and fall off.
▶
Carefully pull on the medical device before each treatment to ensure that it
is securely locked onto the motor coupling.
NOTICE
Removing and attaching the medical device while the drive motor is
rotating.
Damage to the driver.
▶
Never attach or remove the medical device while the drive motor is rotat-
ing.
NOTICE
Pressing the foot switch while attaching or detaching the medical
device.
Property damage to the medical device.
▶
Do not connect or remove the medical device while pressing the foot
switch.
▶
Lightly spray O-rings on the motor coupling with KaVo Spray.
▶
Place the medical device on the motor coupling and lock it into place.
▶
Pull on the medical device to make sure that it is securely affixed to the
coupling.
5.2 Removing the medical device
▶
Unlock the medical device from the motor coupling by twisting it slightly
and then pulling it along its axis.
5.3 Inserting the dental bur or diamond grinder
Note
Only use root canal instruments with shafts that comply with ISO 1797-1
type 1 and ISO 3630-1:
- Shaft diameter: Ø 2.334 to Ø 2.350 mm
- Shaft clamping length: at least 12 mm