ALJ/ALS/PLJ/PLS-BA-e-1135
13
2 Explanation of graphical symbols
Note: follow the directions included
in the attached document or
Follow the Operating Instructions
3 Basic Information (General)
3.1 Proper use
The balance you purchased is intended to determine the weighing value of material
to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the
material to be weighed is manually and carefully placed in the centre of the weighing
plate. As soon as a stable weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use balance for dynamic weighing. In the event that small quantities are
removed or added to the material to be weighed, incorrect weighing results can be
displayed due to the “stability compensation“ in the balance. (Example: Slowly
draining fluids from a container on the balance.)
Do not leave permanent load on the weighing plate. This may damage the
measuring system. Impacts and overloading exceeding the stated maximum load
(max) of the balance, minus a possibly existing tare load, must be strictly avoided.
Balance may be damage by this. Never operate balance in explosive environment.
The serial version is not explosion protected.
The structure of the balance may not be modified. This may lead to incorrect
weighing results, safety-related faults and destruction of the balance.
The balance may only be used according to the described conditions. Other areas of
use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
•
Our conditions in the operation manual are ignored
•
The appliance is used outside the described uses
•
The appliance is modified or opened
•
Mechanical damage or damage by media, liquids, natural wear and tear
•
The appliance is improperly set up or incorrectly electrically connected
•
The measuring system is overloaded
3.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the
balance and, if applicable, the testing weight, must be checked regularly. The
responsible user must define a suitable interval as well as type and scope of this
test. Information is available on KERN’s home page (www.kern-sohn.com with
regard to the monitoring of balance test substances and the test weights required for
this. In KERN’s accredited DKD calibration laboratory test weights and balances may
be calibrated (return to the national standard) fast and at moderate cost.