MWA-M / MWA-PM-BA-e-1614
33
12 Menu
Access to service menu „tCH“ is locked in verified balances.
To disable the access lock, destroy the seal and actuate the
adjustment switch. For position of adjustment switch, see chap. 17.
Attention
:
After destruction of the seal the weighing system must be re-verified
by an authorised agency and a new verification wire/seal mark fitted
before it can be reused for applications subject to verification.
12.1 Navigation in the menu
Call up menu
In weighing mode, press
and the first function
[F1
oFF]
will be displayed.
Select function
With help of
, the individual functions can be
selected one after the other.
Change settings
Confirm selected function by
. The current setting
will be displayed.
Select desired setting by
and confirm with
,
the balance returns to the menu.
Exit menu/
Return to weighing
mode
Press
, the balance will return to weighing mode.