MWA-M / MWA-PM-BA-e-1614
27
11.1.1 Weighing with wheelchair
Place wheelchair and patient in the centre of the weighing balance.
Engage brakes on wheelchair
Do not leave the patient unattended
Wait until patient is sitting still and read weighing value 1.
Loosen the brakes and carefully pull off the transportation stretcher/wheelchair
with patient
Then weigh the wheelchair without the patient and deduct the result from
weighing value 1, from this follows the weight for the patient.
11.2 Taring
The tare weight of any preloads can be deducted by pressing a button so that the
actual weight of the person is displayed in subsequent weighings.
(example)
Put an object (such as a towel, support or empty wheelchair)
on the weighing pan.
Press
, the zero display appears.
„NET” is shown at the bottom on the left.
(example)
Allow the person to step onto the centre of the weighing
platform.
Wait until the standstill display „STABLE“ appears, then read
the weighing result.
When the balance is unloaded the saved taring value is
displayed with negative sign.
To delete the stored tare value, release scales and press
.