MWA-M / MWA-PM-BA-e-1614
23
10.6 Battery operation is possible by obtaining an optional battery power pack
Open the battery compartment cover (1) at the base of the display unit and insert the
rechargeable battery. Charge the battery for at least 12 hours before initial use.
The appearance of the symbol
in the weight display indicates that the battery
is almost exhausted. The weighing scale will remain ready for operation for a few
more minutes before switching off in order to save battery. Load rechargeable
battery.
Voltage has dropped below prescribed minimum.
Rechargeable battery very low.
Rechargeable battery completely reloaded
Right underneath the display there is a LED with the symbol
.
If the LED lights green, the rechargeable battery is fully charged. If it is lighting blue,
it is being charged.
If the balance is not used for a longer time, take out the rechargeable battery and
store it separately. Leaking liquid could damage the balance.