MLS-BA-defsi-0713
132
Fault messages
No.
Description
Error of control sum
1.1
Error data transfer
Error A/D
1.2
Error A/D converter
A/D Null
2.3
Error A/D converter
A/D Full
2.4
Error A/D converter
Tara/Zero outside the range
2.5
Outside the weighing range
Tara outside the range
2.6
Outside taring range
Result > 10% Max
2.7
Result > 10%
Result > 4% Max
2.8
Load on weighing plate too large
The difference > 1% Max
2.9
Difference cal weight to stored
/momentaneous cal. weight > 1%
The sample mass < 1g
2.10
Sample < 1g
The sample mass outsider
the range
2.11
Sample outside the adjusted area
Outside range
3.1
Value of parameter outside
Faulty value
3.2
Value of parameter outside
Incorrect password
3.4
Wrong password
4.1
Parity error
4.2
Table error
4.3
Suspended transmission CTS
4.4
Suspended
transmissionXOFF
4.5
Error data transfer
Incorrect date
5.1
Wrong date
Overcrossed time
6.1
Timeout
Switch the balance off if other error messages should appear and then switch on
again. If error message persists, inform manufacturer.