MLS-BA-defsi-0713
89
2
.
Carrying out with AUTO-Temperature calibration set MLS-A01
•
Temperature calibration
If you only need to measure the temperature within the MLS (only calibration),
then carry out the procedure as described under chapter 6 "Temperature
adjustment”, without connecting the RS 232 interface.
•
Temperature adjustment
The temperature will be measured at two points and will be automatically
adjusted.
To adjust the temperature the following settings are required on the MLS-A01:
Switch on the device with the
key. The device
carries out a self-test. All symbols are shown in the
display:
Note:
When calibrating the temperature, the automatic switch-off function must be
deactivated as follows:
Keep pressing the
key until the time is
displayed
25.2
°C
z
time
temp
alarm
L
..26
08
°14
time
temp
alarm
Keep pressing the
key until
A.OFF
and the
current switch-off time is displayed.
To deactivate the function use the
or
keys
to set the switch-off time to
1
.
A.OFF
°1
z
time
temp
alarm
CAL
time
temp
ppm ppt
uMM/l
mV mS
pH uS
%mg/l
REL
alarm
-
8
8
8
8
1
.
.
.
.
.
.
E1E2E3
k
8
8
8
8
1
.
.
.
.
.
.
C
%
cm
-1
P6
P4
P3
P2
P1