MLS-BA-defsi-0713
111
9.3 Select turn-off criterion
The user can determine which turn-off criteria are to be used
Operator
Indication
Ö
Actuate -key, use the cursor (
►
) to
select the desired setting.
The following turn-off criteria are available:
Ö
Turn-off when weighing value changes <
1mg within 120s
Ö
Turn-off when weighing value changes
< 1mg within 240s
Ö
Turn-off when weighing value changes
< 1mg in freely defined intervals
Ö
Turn-off when weighing value changes
Percent range between 0,001% and 9,999%
Ö
Turn-off when weighing value changes
< 1mg in freely defined intervals
Ö
Turn-off drying process manually
Ö
Turn-off drying process time-controlled
Push
-button
Actuate
-key twice now
Use the
-key to save the setting; with the
-key the old setting remains valid.