KERN MBC M Скачать руководство пользователя страница 6

 

 

MBC-M-SH-e-1310 

3.7  Trouble Shooting 

 
Problems 

Possible cause 

Common Solutions 

Display is blank. 
No self test 

Mains power is turned off. 
Power supply faulty or not 
plugged. Internal battery is 
not charged. On/Off switch 
problem 

Check power is getting inside the scale and 
on/off switch is working. 
Verify the voltages, which is on the power 
labels. 

Blank display after 
self test 

Pan not installed. 
Unstable weight, load cell 
damaged 

Check the pans are installed correctly. Try 
to turning on again. 
 

 
OL or ------- 

 
Maximum capacity 
exceeded. Load cell or 
mechanics damaged. 
Power supply faulty 
 

 
Check the platform is installed correctly. 
Try to turn on the scale again. Do the 
calibration again 

 
------- or NULL 
displayed 

Weight is on the platform is 
below permissible limit.  
Pan not installed correctly. 
Power supply faulty. 
Load cell or mechanism 
faulty 
 

Check the platform is installed correctly. 
Try to turn on the scale again. 
Do the calibration again 

Display is unstable 

Goods touching 
somewhere. Air variation or 
any vibrations. 
Temperature changed . 
Load cell or connections 
faulty. 
Power supply faulty  

Check the scale is in acceptable location. 
Check the connecters and load cell. 
Check the power supply and battery 

Weight value 
incorrect 

Calibration error.  
Platform of load cell 
touching somewhere. 
Wrong weighing unit  

Use accurate weight for to do the 
calibration  
Check the pan and load cell is installed 
proper and touching. 
Check the parameter settings. 
Check the load cell and connecters 

Can not use full 
capacity 

Over load protection 
stoppers or transport locks 
are not removed. 
Parameters are set 
incorrectly.  
AD problem. 
Load cell or mechanism 
damaged 
 

Check the stoppers and locks under the 
platform. 
Check the weighing unit and parameter 
settings. 
Check the load cell. 

Platform Corner 
Weight different 

Over load protection 
stoppers or transport locks 
are not removed. 
Load cell or mechanism 
damaged 

Check the stoppers and locks under the 
platform. 
Use accurate weight for to do the 
calibration  
Check the load cell. 

Battery not charging 

Mains voltage problem 
Charging circuit problem 
Battery Problem 

Check the mains and adaptor. 
Check the battery. 
Check the charging circuit 

 
 

Содержание MBC M

Страница 1: ...n GmbH Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Service Manual Baby scale KERN MBC_M Version 1 0 05 2013 GB MBC M SH...

Страница 2: ...nts 1 Basic Information 3 2 Introdution 3 3 MAINTENANCE 4 3 1 General 4 3 2 Error Codes 4 3 3 Determine the Problem 4 3 4 Check the Load cell 4 3 5 Load Cell Connections 5 3 6 Checking the different V...

Страница 3: ...t 100 compliance with the type approval A violation of this specification will result in a loss of the type approval After successful repair the balance will have to be reverified before it can be use...

Страница 4: ...he platform Overload when zeroing the scale Improper calibration Load cell problem PCB problem Err 6 A D Count out of the range Platform not installed Load cell problem PCB problem 3 3 Determine the P...

Страница 5: ...nnecter voltage Compare the value with the name plate on the scale PCB Input Voltage Check the PCB input power connecter voltage Voltage must be minimum 9VDC in to the pin AD Check Battery Voltage and...

Страница 6: ...variation or any vibrations Temperature changed Load cell or connections faulty Power supply faulty Check the scale is in acceptable location Check the connecters and load cell Check the power supply...

Страница 7: ...MBC M SH e 1310 7 4 DRAWING 4 1 MBC HM...

Страница 8: ...8 MBC M SH e 1310 4 2 MBC oHM...

Страница 9: ...4001 Key panel 20kg 27 19 RMBC 4002 Battery cover 37 29 RMCC 8001 MBC A04 MBC A04 Footstand 1 2 48 49 41 42 RMBC 8001 Level bubble 26 18 RMPE 8003 Adapter socket 40 34 RMCC 5001 Plastic scale pan 1 1...