3.5  Load Cell Connections 

 

Signal    +    

Green 

Supply   +    

Red

 

 
 
Signal    -    

Black

 

 
 
Supply   -    

White

 

Shield 
 
Measuring  Points 

Resistance 

Red (+ Exc) to White ( 

–Exc)  

420 ±20Ω 

Green (+Sig) to Black ( 

–Sig)  

350Ω ±5Ω 

 

3.6  Checking the different Voltages 

If the problem is in the PCB, use a multimeter to check the following voltages 

 

AC Power 

Check the AC power socket out put voltage.  
 

Adaptor Voltage 

Check the adaptor output cable connecter voltage 
Compare the value with the name plate on the scale 
 

PCB Input Voltage  

Check the PCB input power connecter voltage 
Voltage must be minimum 9VDC in to the pin AD+  
 

Check Battery Voltage and Charging Voltage 

Check the Battery Voltage

Voltage must be minimum 6VDC. If below the 6VDC connect the adaptor for charging 
The battery voltage below the 5.5VDC, replace the battery and install new 6V/3.4Ah battery.  
 
Check the Battery Charging Voltage; 
Remove the battery connection terminals (Red and Black) from the battery. Connect the power 
and turn on the scale Voltage into the terminal minimum 6.5VDC 
 

 
 

 

Содержание MBC M

Страница 1: ...n GmbH Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Service Manual Baby scale KERN MBC_M Version 1 0 05 2013 GB MBC M SH...

Страница 2: ...nts 1 Basic Information 3 2 Introdution 3 3 MAINTENANCE 4 3 1 General 4 3 2 Error Codes 4 3 3 Determine the Problem 4 3 4 Check the Load cell 4 3 5 Load Cell Connections 5 3 6 Checking the different V...

Страница 3: ...t 100 compliance with the type approval A violation of this specification will result in a loss of the type approval After successful repair the balance will have to be reverified before it can be use...

Страница 4: ...he platform Overload when zeroing the scale Improper calibration Load cell problem PCB problem Err 6 A D Count out of the range Platform not installed Load cell problem PCB problem 3 3 Determine the P...

Страница 5: ...nnecter voltage Compare the value with the name plate on the scale PCB Input Voltage Check the PCB input power connecter voltage Voltage must be minimum 9VDC in to the pin AD Check Battery Voltage and...

Страница 6: ...variation or any vibrations Temperature changed Load cell or connections faulty Power supply faulty Check the scale is in acceptable location Check the connecters and load cell Check the power supply...

Страница 7: ...MBC M SH e 1310 7 4 DRAWING 4 1 MBC HM...

Страница 8: ...8 MBC M SH e 1310 4 2 MBC oHM...

Страница 9: ...4001 Key panel 20kg 27 19 RMBC 4002 Battery cover 37 29 RMCC 8001 MBC A04 MBC A04 Footstand 1 2 48 49 41 42 RMBC 8001 Level bubble 26 18 RMPE 8003 Adapter socket 40 34 RMCC 5001 Plastic scale pan 1 1...

Отзывы: