KERN MBC M Скачать руководство пользователя страница 4

 

 

MBC-M-SH-e-1310 

3  MAINTENANCE

                                        

 

 

3.1  General  

 

Please clean the scale each week and do calibration every year. 
If the scale does not operate properly, find out the problem as possible. 
Determine whether the problem is constant or alternate. Be aware that problems can be caused by 
mechanical or electrical influences. 
 
Check the following. 

 

Water 

 

Corrosive materials 

 

Vibrations or temperature or wind 

 

Physical damage 

 
 

Check the scale cables for damage, and check all connections and connecters for any loose contact 
or incorrect connection 

 

3.2  Error Codes 

   

 

 

Error Code 
 

Description 

POSSIBLE CAUSES 

Err 4 

Zero range exceeded, due to 

turning on or by pressing 

 

 

Goods on the platform 

 

Overload, when zeroing the 
scale. 

 

Improper calibration 

 

Load cell problem 

 

PCB problem 

Err 6 

A/D Count out of the range 

 

Platform not installed  

 

Load cell problem 

 

PCB problem 

 

3.3  Determine the Problem 

Determine whether the problem is in the PCB or the Load Cell 
 

 

Disconnect the power supply from the scale, and disconnect the load cell connection from the 
PCB 

 

Reapply power and test the PCB 

 

If problem goes away, its source is probably in the Load cell. Check the wiring, connecter, 
load cell and mechanical components of the load cell. 

 

If problem persists, its source is probably in the PCB. Check the PCB voltages, connecters, 
cables and function programs 

 

 

3.4  Check the Load cell 

 

 

Remove power from the system, and disconnect the PCB  from the Load cell 

 

Make sure all leads are connected and correctly.  

 

Check load cell for proper input and output resistances 

 
 
 
 

Содержание MBC M

Страница 1: ...n GmbH Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Service Manual Baby scale KERN MBC_M Version 1 0 05 2013 GB MBC M SH...

Страница 2: ...nts 1 Basic Information 3 2 Introdution 3 3 MAINTENANCE 4 3 1 General 4 3 2 Error Codes 4 3 3 Determine the Problem 4 3 4 Check the Load cell 4 3 5 Load Cell Connections 5 3 6 Checking the different V...

Страница 3: ...t 100 compliance with the type approval A violation of this specification will result in a loss of the type approval After successful repair the balance will have to be reverified before it can be use...

Страница 4: ...he platform Overload when zeroing the scale Improper calibration Load cell problem PCB problem Err 6 A D Count out of the range Platform not installed Load cell problem PCB problem 3 3 Determine the P...

Страница 5: ...nnecter voltage Compare the value with the name plate on the scale PCB Input Voltage Check the PCB input power connecter voltage Voltage must be minimum 9VDC in to the pin AD Check Battery Voltage and...

Страница 6: ...variation or any vibrations Temperature changed Load cell or connections faulty Power supply faulty Check the scale is in acceptable location Check the connecters and load cell Check the power supply...

Страница 7: ...MBC M SH e 1310 7 4 DRAWING 4 1 MBC HM...

Страница 8: ...8 MBC M SH e 1310 4 2 MBC oHM...

Страница 9: ...4001 Key panel 20kg 27 19 RMBC 4002 Battery cover 37 29 RMCC 8001 MBC A04 MBC A04 Footstand 1 2 48 49 41 42 RMBC 8001 Level bubble 26 18 RMPE 8003 Adapter socket 40 34 RMCC 5001 Plastic scale pan 1 1...

Отзывы: