2 Technical 

Data 

 

KERN 

MAP 80K1 

Capacity  

80 kg 

(blue) 

40 kg 

(green) 

20 kg 

(red) 

 

Readability (d) 

0.1 kg / 0.2 lb 

Units kg, 

lb 

Auto Off  

After 1 min without change of load  

Electric Supply 

1 x CR2450 

Operating temperature 

+ 5°C … + 35°C 

Storage temperature 

- 20°C … + 60°C 

Humidity of air 

max. 80 % (not condensing) 

Dimensions fully mounted  
(W x D x H) mm 

212 x 55 x 102 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание MAP 80K1

Страница 1: ...mbH Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Service manual Hand grip dynamometer KERN MAP Version 1 0 3 2010 GB MAP...

Страница 2: ...grip dynamometer Table of Contents 1 Basic Information 3 2 Technical Data 4 3 Overview of Display Keys 5 4 Error Message Indication 6 5 Trouble Shooting 7 6 Calibration Procedure 8 7 Zero Count Adjus...

Страница 3: ...tory for repair work After repair original conditions of the device have to be restored Only original spare parts should be used Das Ger t darf nur von geschultem oder beruflich ausgebildetem Fachpers...

Страница 4: ...en 20 kg red Readability d 0 1 kg 0 2 lb Units kg lb Auto Off After 1 min without change of load Electric Supply 1 x CR2450 Operating temperature 5 C 35 C Storage temperature 20 C 60 C Humidity of air...

Страница 5: ...s 1 Stability display 2 Battery status 3 Operation modus 4 Percentage display 5 Weighing unit lb 6 Weighing unit kg Keyboard overview Button Function Zero Switch on off zeroing Mode Select operating m...

Страница 6: ...MAP SH e 1010 6 4 Error Message Indication Capacity of batteries exhausted Underload set to zero by pressing zero key Overload Preload error EEPROM error...

Страница 7: ...MAP SH e 1010 7 5 Trouble Shooting...

Страница 8: ...g 1933 3 Press the zero key once The display appears 0 4 Load calibration weight 40kg The display shows the internal span count e g 1146 5 Press the zero key once to start the calibration Shortly aft...

Страница 9: ...variable resistor on R1 to adjust the zero count If zero count is above 3000 use a variable resistor on R2 to adjust the zero count After determination of the required resistor solder it on the main b...

Страница 10: ...MAP SH e 1010 10 8 PCB Layout Load cell Battery connector...

Страница 11: ...MAP SH e 1010 11 9 Display Format...

Страница 12: ...MAP SH e 1010 12 10 Assembly Drawing...

Отзывы: