13.2 How to connect the platform
Disconnect the display unit from the power supply.
Connect load cell cable to the connecting plug. Please see diagram below for plug
allocation.
4-(5-)wire weighing cells
6-(7-)wire weighing cells
Pin 1
SIG+
Pin 1
SIG+
Pin 2, 3
EXC-
Pin 2
SEN-
Pin 4, 5
EXC+
Pin 3
EXC-
Pin 6, 7, 8
Screen
Pin 4
SEN+
Pin 9
SIG-
Pin 5
EXC+
Pin 6, 7, 8
Screen
Pin
9
SIG-
Connect the plug to the display unit.