6.7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each display unit with connected weighing plate must be coordinated - in compliance
with the underlying physical weighing principle - to the existing acceleration due to
gravity at its place of location (only if the weighing system has not already been
adjusted to the location in the factory). This adjustment process must be carried out
for the first commissioning, after each change of location as well as in case of
fluctuating environment temperature. To receive accurate measuring values it is also
recommended to adjust the display unit periodically in weighing operation.
+
Carry out adjustment, see chapter 14.2
6.8 Verification
General introduction
:
According to EU directive 90/384/EEC or 2009/23EG balances must be officially
verified if they are used as follows (legally controlled area):
a) For commercial transactions if the price of goods is determined by weighing.
b) For the production of medicines in pharmacies as well as for analyses in the
medical and pharmaceutical laboratory.
c) For official purposes
d) For manufacturing final packages
In cases of doubt, please contact your local trade in standard.
Verification notes
:
An EU type approval exists for the appliance described in its technical data as
verifiable. If the appliance is used where obligation to verify exists as described
above, it must be verified and re-verified at regular intervals.
Re-verification of an appliance is carried out according to the respective national
regulations.
Normally t
he validity for verification in Germany is e.g. 2 years.
The legal regulation of the country where the balance is used must be observed!
Verification of the device is invalid without the seals
.
The seal marks attached on verified appliances point out that the appliance
may only be opened and serviced by trained and authorised specialist staff. If
the seal mark is destroyed, verification looses its validity. Please observe all
national laws and legal regulations. In Germany a re-verification will be
necessary.
14
KXE-TM-BA_IA-e-1110