3.4  Monitoring of Test Resources 

In the framework of quality assurance the measuring-related properties of the display 
unit and, if applicable, the testing weight, must be checked regularly. The responsible 
user must define a suitable interval as well as type and scope of this test. Information 
is available on KERN’s home page (

www.kern-sohn.com

 with regard to the 

monitoring of display units’ test substances and the test weights required for this. In 
KERN’s accredited DKD calibration laboratory test weights and display units may be 
calibrated (return to the national standard) fast and at moderate cost. 

4  Basic Safety Precautions 

4.1  Pay attention to the instructions in the Operation Manual 

Carefully read this operation manual before setup and commissioning, even if you 
are already familiar with KERN balances. 

4.2  Personnel training 

The appliance may only be operated and maintained by trained personnel. 

5  Transport and storage 

5.1  Testing upon acceptance 

When receiving the appliance, please check packaging immediately, and the 
appliance itself when unpacking for possible visible damage. 

5.2  Packaging / return transport 

 

 

Ö

 Keep all parts of the original packaging for a possibly required 

return. 

Ö

 Only use original packaging for returning. 

Ö

 Prior to dispatch disconnect all cables and remove 

loose/mobile parts.  

Ö

 Reattach possibly supplied transport securing devices.  

Ö

 Secure all parts such as the glass wind screen, the weighing 

platform, power unit etc. against shifting and damage. 

 

 
 
 

KXE-TM-BA_IA-e-1110   

11 

Содержание KXE-TM

Страница 1: ...D 72336 Balingen E Mail info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating and Installation Instructions Display Unit KERN KXE TM Version 1 0 06 201...

Страница 2: ...arranty 10 3 4 Monitoring of Test Resources 11 4 Basic Safety Precautions 11 4 1 Pay attention to the instructions in the Operation Manual 11 4 2 Personnel training 11 5 Transport and storage 11 5 1 T...

Страница 3: ...e menu 27 10 2 Display higher resolution 28 10 3 Peak value function 29 11 Service maintenance disposal 30 11 1 Clean 30 11 2 Service maintenance 30 11 3 Disposal 30 11 4 Error messages 31 12 Instant...

Страница 4: ...nges max 2 Weighing Units kg t Divisions 1 2 5 10 n Display LCD 25 mm digits with back lighting DMS weighing cells Min 87 Max 1000 Input voltage 230 V 50 60 Hz Electric Supply Mains adapter secondary...

Страница 5: ...KXE TM BA_IA e 1110 5 Dimensions 1 With wall support 2 Without wall support...

Страница 6: ...view 1 7 2 8 3 4 5 6 1 Weight display 2 Battery compartment 3 RS 232 disabled 4 Connection power supply mains adapter 5 Adjustment switch 6 Connection platform 7 Fastening screws 8 Wall bracket 6 KXE...

Страница 7: ...Turn on off Quit Back to menu weighing mode Zeroing Taring Change between gross and net weight Functional button allocation see chap 9 1 FNC 02 Functional button allocation see chap 9 1 FNC 03 Confirm...

Страница 8: ...t selection to the left Digit selection to the right Save setting Reject setting 2 2 Function keys The following functions can be allocated to the buttons or in menu see chap 9 FNC 02 FNC04 0 Printing...

Страница 9: ...are value is stored Instable weighing value M Data in summation memory GROSS that the displayed weighing value is a gross weight value PRE TARE that the displayed weight value is a PRE TARE value Reso...

Страница 10: ...damage the measuring system Impacts and overloading exceeding the stated maximum load max of the weighing plate minus a possibly existing tare load must be strictly avoided Both the weighing plate and...

Страница 11: ...o the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning even if you are already familiar with KERN balances 4 2 Personnel training The appliance...

Страница 12: ...nit to extreme dampness for longer periods of time Non permitted condensation condensation of air humidity on the appliance may occur if a cold appliance is taken to a considerably warmer environment...

Страница 13: ...pplied via the external mains adapter The stated voltage value must be the same as the local voltage Only use original KERN mains adapters Using other makes requires consent by KERN 6 6 Battery operat...

Страница 14: ...etermined by weighing b For the production of medicines in pharmacies as well as for analyses in the medical and pharmaceutical laboratory c For official purposes d For manufacturing final packages In...

Страница 15: ...Position of seals and adjusting switch KXE TM BA_IA e 1110 15...

Страница 16: ...ss for 3 sec the display disappears 7 3 Zeroing Resetting to zero corrects the influence of light soiling on the weighing plate To unload the weighing system Press and zero display as well as indicato...

Страница 17: ...o display the indicator T above TARE and NET will appear The weight of the container is now internally saved Weigh the material the net weight will be indicated After removing the weighing container t...

Страница 18: ...p 9 Functional button Setting FNC 02 6 or FNC 03 6 or FNC 04 6 Press the functional button PT followed by 0000 00 will be displayed The active digit is flashing Enter the known tare weight e g 2 kg an...

Страница 19: ...al buttons see chap 9 Functional button Setting Description FNC 02 4 Display of all add up data Number weighing processes Total Weighing mode FNC 03 2 M Totalization commutation add up mode weighing m...

Страница 20: ...is added into the summation memory Add more weighed goods as described before You can repeat this process until the capacity of the weighing system is exhausted Display switch over weighing data Pres...

Страница 21: ...e total added memory With display Number of weighings or Total weight press at the same time The data in the summation memory are deleted Weighing mode Indicator T over M disappears KXE TM BA_IA e 111...

Страница 22: ...9 Function menu Call up menu In weighing mode press and at the same time the functional menu 01 FnC is displayed Switch over between and not documented using and 22 KXE TM BA_IA e 1110...

Страница 23: ...quently using FNC 00 FNC 08 see chap 9 1 Confirm selected function by the current setting is displayed Switch into the available settings using and Confirm selection by the next function is displayed...

Страница 24: ...0 none 1 20 times s 2 10 times s 3 5 times s FNC 01 Update of display 4 1 times s 0 Print 1 Switch over weighing unit 2 M adding up printing 3 MC delete sum memory 4 Display switch over in add up mode...

Страница 25: ...int 1 Switch over weighing unit 2 M adding up printing 3 MC delete sum memory 4 Display switch over in add up mode Number weighing processes Total Weighing mode 5 HR conversion into higher resolution...

Страница 26: ...1 Mode 1 see chap 11 3 FNC 07 Peak value function 2 Mode 2 see chap 11 3 FNC 08 Auto power Off Automatic power switch off 1 60 min The display unit is automatically switched off within the preset time...

Страница 27: ...L is displayed Press and keep it pressed until End appears Back to weighing mode actuate adjustment switch Reset all parameters to allowed settings for verified weighing systems to switch on the unit...

Страница 28: ...etting must be allocated to a functional button see chap 9 Functional button Setting FNC 02 5 or FNC 03 5 or FNC 04 5 When pressing the functional button the weight value is displayed approx 5 seconds...

Страница 29: ...alue of 10 d the highest load value will be displayed The peak value remains in the display until it will be deleted using Then the balance is ready for further measurements Display Mode 2 Menu settin...

Страница 30: ...roxide solution acetic hydrochloric sulphuric or citric acid Do not use metal brushes or cleaning sponges of steel wool as this causes superficial corrosion 11 2 Service maintenance The appliance may...

Страница 31: ...he A D changer range Err 2 Weighing value Zero range Err 6 Internal solution 0 15 V d range Err Wrong password E 1 Internal value zero range E 2 Internal value zero range Should other error messages o...

Страница 32: ...tteries are inserted incorrectly or empty No rechargeable batteries inserted The displayed weight is permanently changing Draught air movement Table floor vibrations Weighing plate has contact with ot...

Страница 33: ...iest load to be weighed Preload This corresponds to the total weight of all parts that are to be placed on the weighing cell such as upper part of platform weighing pan etc Total zero setting range Th...

Страница 34: ...the connecting plug Please see diagram below for plug allocation 4 5 wire weighing cells 6 7 wire weighing cells Pin 1 SIG Pin 1 SIG Pin 2 3 EXC Pin 2 SEN Pin 4 5 EXC Pin 3 EXC Pin 6 7 8 Screen Pin 4...

Страница 35: ...hing system must be re verified by an authorised agency and a new verification wire seal mark fitted before it can be reused for applications subject to verification Call up menu Actuate adjustment sw...

Страница 36: ...ect the individual functions one by one Confirm selected function such as CSP 04 with the current setting is displayed Switch into the available settings using and Confirm selection by the next functi...

Страница 37: ...on decimal point d 0 0000 4 decimal places CSP 01 Balance capacity Max 999999 000000 The possible capacities and factory settings depend on the connected weighing bridge 1 2 5 10 20 CSP 02 Readability...

Страница 38: ...d 1 25 d 2 s 1 5 d 1 5 d 2 s 1 75 d 1 75 d 2 s 2 d 2 d 2 s CSP 04 Zero Tracking no Deactivated Digits time 0 25 d 0 25 d 1 s 0 5 d 0 5 d 1 s 0 75 d 0 75 d 1 s 1 d 1 d 1 s 1 25 d 1 25 d 2 s 1 5 d 1 5 d...

Страница 39: ...weight Info about test weights can be found on the Internet at http www kern sohn com Observe stable environmental conditions Stabilisation requires a certain warm up time Actuate adjustment switch S...

Страница 40: ...ed followed by the currently set weight value for the adjustment weight Either use the displayed adjustment weight or change it with and numerical input see chapter 2 1 1 the active digit is flashing...

Страница 41: ...owledge with The currently set password is displayed the active digit flashes In 0000 factory setting no password is deposited Enter the desired password e g 1234 Numeric input see chap 2 1 1 and conf...

Страница 42: ...ollowed by Enter the password e g 1234 Numeric input see chap 2 1 1 and confirm by If the password is entered correctly settings in the menu may be carried out If a wrong password is entered Err is di...

Страница 43: ...to select the individual functions one by one Confirm selected function e g CFn 05 with the current setting is displayed Switch into the available settings using and Confirm selection by the next func...

Страница 44: ...n CFN 01 Set to zero during start up 1 Off CFN 02 Zeroing range 2 always use this setting for verified weighing systems CFN 03 Adaptation of weighing speed factory setting 6 0 9 fast slow 0 Calm and s...

Страница 45: ...Animal weighing mode 2 Not documented CFN 06 Not documented 10 CFN 07 Not documented 2 Always use this setting 0 Multi interval balance CFN 08 Balance Type 1 Multi range balance CFN 09 Setting multi r...

Отзывы: