12
MBB-M-BA-e-1617
Guidance and manufacturer’s declaration-electromagnetic immunity
The MBB15K2DM is intended for use in the electromagnetic environment specified below.
The customer or the user of the MBB15K2DM should assure that is used in such and environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Radiated RF IEC
61000-4-3
3 Vrms
150 KHz to 80 MHz
3 V/m 80MHz to 2,5
GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the MBB15K2DM;including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d = 1,2
√
P
d = 1,2
√
P
80MHz to 800 MHz
d = 2,3
√
P
800MHz to 2,5 GHz
Where
P
is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in
each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following symbo
l:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the MBB15K2DM is used exceeds the applicable RF
compliance level above, the MBB15K2DM should be observed to verify normal operation. If abnormal performance is
observed, additional measures my be necessary, such as re-orienting or relocating the MBB15K2DM.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be les than 3 V/m.
Содержание KERN MBB-M series
Страница 41: ...MBB M BA d 1617 40 Die Auswahl erfolgt mit der HOLD und TARE Taste Werkseinstellung ...
Страница 82: ...40 MBB M BA e 1617 Press the HOLD and TARE buttons to make the selection default setting ...
Страница 85: ......
Страница 107: ...MBB M BA f 1617 23 Branchement CN 3 de pile rechargeable Branchement CN 4 de piles AA x 6 ...
Страница 125: ...MBB M BA f 1617 41 La sélection s opère sur la touche HOLD et TARE réglé en usine ...