34
HFD-BA-e-1820
10. Error messages
Error
message
Description
Possible causes / Elimination
Err 1
Wrong date
Enter date in format "yy;mm;dd", see
Err 2
Wrong time
Enter time in format
“hh;mm;ss“
, see
Err 4
Error during zero setting
Zero range exceeded
Check if the balance is not loaded
Err 5
Keyboard error
Improper operation of the balance
Err 6
Value outside the A/D changer
range
Damaged weighing cell
Damaged electronics
Err 7
Error "Determination of
percentage"
Increase the value to >0.5 d
Err 8
Incorrect adjusting weight
Check value for adjustment weight, see
chap. 1
Err 9
The displayed weight is
permanently changing
Draught/air movement
Table/floor vibrations
Weighing pan has contact with other
objects.
Err 10
No WLAN connection
Check menu setting
"F5 com
mode
wifi"
Err 11
Error "Communication log"
Check the communication settings
Err 12
Error "Totalizing"
Number of totalizing operations > 99
Capacity of balance exhausted
Err 15
Error "Gravitational constant"
Value out of range 09.xx -1.0xx
Err 17
"Taring" error
Taring range exceed or not achieved
Err 19
Unable to initialise zero point
Measuring cell defective / overloaded
Object on weighing pan / contact
Main board defective
Adjustment required
- - ol - -
Maximum load exceeded
Reduce load
Check whether the balance has been
damaged
--lo--
Underload
Negative weight, check platform and
restart or adjust.
Fai l h / fai l l
/ fai l
Adjustment error
Check value for adjustment weight, see
chap. 1
Repeat adjustment process
Ba lo / lo ba
Capacity of rechargeable battery
exhausted
Recharge battery
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
Содержание HFD 10T-3
Страница 8: ...7 HFD BA d 1820 1 1 Abmessungen mm ...
Страница 34: ...33 HFD BA d 1820 Justierschalter und Eichdraht 1 Justierschalter 2 Eichdraht ...
Страница 41: ...HFD BA d 1820 40 ...
Страница 44: ...HFD BA d 1820 43 Kranwaage Modell Serien Nr Teil Maßnahme Datum Name Unterschrift ...
Страница 50: ...HFD BA e 1820 7 1 1 Dimensions mm ...
Страница 67: ...24 HFD BA e 1820 Do not push or pull Do not pull the hook horizontally Do not use multiple suspensions ...
Страница 76: ...HFD BA e 1820 33 Adjustment button and verification seal wire 1 Adjustment button 2 verification seal wire ...
Страница 82: ...HFD BA e 1820 39 ...
Страница 85: ...HFD BA e 1820 42 Crane scales Model Serial no Part Action Date Name Signature ...
Страница 91: ...HFD BA f 1820 7 1 1 Dimensions mm ...
Страница 123: ...HFD BA f 1811 39 ...
Страница 126: ...HFD BA f 1811 42 Balance à grue Modèle N de série Pièce constitutive Mesure Date Nom Signature ...