22
GAB-N-BA-e-1817
Operation
Display
How to carry out linearization:
Switch on the balance and during the
selftest press the
and the
button at the same time.
„LoAd 0“ followed by LoAd 1“ will be
displayed.
Place the second adjustment weight.
After the apparition of the stability display,
„LoAd 2“ will be displayed
Put on the third adjustment weight
After the apparition of the stability display,
„LoAd 3“ will be displayed
Put on the forth adjustment weight
(max load)
After the apparition of the stability display,
„LoAd 2“ will be displayed
Third adjustment weight
After the apparition of the stability display,
„LoAd 1“ will be displayed
Second adjustment weight
After the apparition of the stability display,
„LoAd 0“ will be displayed
First adjustment weight (weighing plate
empty)
After successful linearization the balance
automatically returns to weighing mode.
In case of an adjustment error or incorrect adjusting weight the display will show an
error message; repeat linearization process.