9 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect from
power supply. The weighing process must then be restarted from the beginning.
Help:
Fault
Possible cause
The displayed weight does not glow.
•
The balance is not switched on.
•
The mains supply connection has been inter-
rupted (mains cable not plugged in/faulty).
•
Power supply interrupted.
The displayed weight is permanently
changing
•
Draught/air movement
•
Table/floor vibrations
•
Weighing plate has contact with other objects.
•
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
The weighing value is obviously
•
The display of the balance is not at zero
wrong
•
Adjustment is no longer correct.
•
Great fluctuations in temperature.
•
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
Should other error messages occur, switch balance off and then on again. If the error mes-
sage remains inform manufacturer.