17
OPE-1-BA-e-1610
6
Changing the bulb
Halogen
Before changing the bulb the device must be switched off and unplugged.
To change the bulb, tip the device carefully to the back or side. When doing this,
please make sure that all microscope components are firmly fixed. The bulb holder is
on the underside of the device. It can be opened by undoing the screws on the holder
(see illustration)
. The defective LED module can be removed by loosening the two
screws fixing the module and unraveling the connection point of its cable. Now the
new module has to be mounted in the same why as the original one. After the bulb
holder has been replaced in the underside of the device and the screws replaced, the
bulb replacement procedure is complete.
7
Changing the fuse
The fuse housing is on the rear of the microscope below the mains power supply
socket. With the device switched off and unplugged, you can pull out the housing.
When doing this, it is helpful to use a screwdriver or similar tool. The defective fuse
can be removed from its housing and be replaced with a new one.
After that, you just need to insert the fuse housing back into the insertion point below
the mains power supply socket.
Содержание OPE 118
Страница 2: ......
Страница 7: ...OPE 1 BA e 1610 6 2 Nomenclature...
Страница 9: ...OPE 1 BA e 1610 8...