background image

13 

 

OPE-1-BA-e-1610 

1.  Bring the 10x objective into beam path. 

 

2.  Assure, that one eyepiece with scale is attached to (one of) the tube 

connector(s). 
 

3.  Locate an appropriate specimen slide onto the table. 

This slide should preferably be equipped with a micro reticule. 

It would be also possible to use an object which includes plenty of single dots and for which 
one of those dots has such a size, so that it aligns with the centre cross point of the scale, 
visible inside of the eyepiece(s).

 

 

4.  Locate the specimen slide to that point, that, when observing through the 

eyepiece(s), the centre of the reticule is on the centre of the eyepiece scale. 
 

5.  Assure, that the fixing screw of the table is loosened, in order to be able to 

revolve the table. 

If the table is not or heavy to revolve, even though the fixing screw is loosened, this serves as 
an evidence for a significant decentration of the table.

 

 

6.  If the table is perfectly centre-adjusted, you will note that, during a complete 

rotation of the table, both centres stay always aligned to each other.  
In this case, the procedure would be finished at that point. 
 

7.  If the table is not centre-adjusted, you will note that the centre of the reticule 

moves, directly after the beginning of the rotation of the table, away from the 
centre of the eyepiece scale. And it matches again only after the complete 
rotation. 
 

8.  Estimate the centre of the circular motion, which the reticule is doing, and 

move the specimen slide, so that the centre of the reticule matches this 
estimated centre. 
 

9.  Operate the centring screws, so that the centre of the reticule and the centre 

of the eyepiece scale are aligning to each other again. 
 

10. Repeat steps 6 - 9. 

 

 

Содержание OPE 118

Страница 1: ...RN Sohn GmbH Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Polarisation microscope KERN OPE 1 OPE 118 V...

Страница 2: ......

Страница 3: ...Features 7 4 Assembly 9 4 1 Polarisation unit Microscope head 9 4 2 Objectives 9 4 3 Eyepieces 9 4 4 Colour filter 10 4 5 Condenser Polariser 10 5 Operation 10 5 1 Getting started 10 5 2 Pre focussing...

Страница 4: ...n additional risk of triggering an electric shock 1 2 Notes on the electrical system Before connecting to a mains power supply you must make sure that you are using the correct input voltage The infor...

Страница 5: ...microscope is not being used you should cover it with the enclosed dust protective cover When doing this the power supply is stopped by switching off at the main switch and unplugging the mains cable...

Страница 6: ...these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms However organi...

Страница 7: ...OPE 1 BA e 1610 6 2 Nomenclature...

Страница 8: ...e pieces Objectives Illumination OPE 118 Finite Monocular WF 10x 18 mm Achromatic 4x 10x 40x 6V 20W Halogen Transmitted Product dimensions 360x200x400 mm Packaging dimensions 440x340x240 mm Net weight...

Страница 9: ...OPE 1 BA e 1610 8...

Страница 10: ...should always make sure that you do not touch the lenses with your bare fingers and that no dust enters the apertures The Bertrand lens and the analyser are inherent parts of the polarization unit The...

Страница 11: ...en you need to remove the condenser Then use the focus dial of the condenser to move the condenser holder to a low position In this way the condenser can be taken off after loosening the three screws...

Страница 12: ...soon as an image is recognisable no matter how sharp then you should only adjust the focus using the fine adjustment knob Adjusting the torque of the coarse and fine adjustment knob Next to the left a...

Страница 13: ...g on its angle position between polariser and analyser For getting ideal results the centre of the rotation axis of the table must be aligned to the centre of the optical beam path The microscopes of...

Страница 14: ...If the table is not or heavy to revolve even though the fixing screw is loosened this serves as an evidence for a significant decentration of the table 6 If the table is perfectly centre adjusted you...

Страница 15: ...objectives you should use the silver ring above the objectives see illustration When rotating the nosepiece you must always make sure that the objective which is about to be positioned in the beam pat...

Страница 16: ...ure diaphragm of the condenser to find the very best compromise between contrast and resolution for the microscopic image For the objective with the lowest magnification the lever of the aperture diap...

Страница 17: ...in the field of view The slide of the Bertrand lens needs to be in the pulled out position for standard polarisation processes It can be moved into the beam path in order to observe the interference...

Страница 18: ...connection point of its cable Now the new module has to be mounted in the same why as the original one After the bulb holder has been replaced in the underside of the device and the screws replaced t...

Страница 19: ...y centred The condenser is too low The field of view is dark or is not correctly illuminated The objective is not positioned correctly on the beam path The selector switch for the beam path is between...

Страница 20: ...Dirt dust on the front lens of the condenser Dirt dust on the object One side of the image is blurred The stage was not correctly fitted The objective is not positioned correctly on the beam path The...

Страница 21: ...materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...

Отзывы: