MST-MHT SERIES 071320
2-1
SECTION 2 - INSTALLATION
2.1
UNPACKING AND INSPECTION
This instrument has been thoroughly inspected and tested prior to packing and is ready for
operation. After careful unpacking, inspect for shipping damage before attempting to operate.
Perform the preliminary operational check as outlined in PAR. 2.7. If any indication of damage is
found, file an immediate claim with the responsible transport service.
NOTE
If the Power Modules are shipped pre-installed in the RA 55H Rack Adapter, each
module is secured to the RA 55H Rack Adapter by two shipping screws. These
screws must be removed from the bottom of the rack adapter, otherwise the mod-
ules cannot be removed from the rack adapter.
2.2
TERMINATIONS AND CONTROLS
a. Internal Calibration Controls: Refer to Figure 2-2 and Table 2-1.
b. Front Panel: Refer to Figure 2-1 and Table 2-2.
c. Rear Panel: Refer to Figure 2-1 and Table 2-3.
2.3
A-C INPUT REQUIREMENT
MST-MHT Power Modules operate from single-phase a-c mains power over the specified volt-
age and frequency ranges without adjustment or modification.
2.4
COOLING
The power transistors and rectifiers in the MST-MHT Power Module are maintained within their
operating temperature range by means of internal heat sink assemblies cooled by internal fans.
PANEL OPENINGS AND THE TOP OF THE CASE MUST BE KEPT CLEAR FROM
OBSTRUCTIONS TO INSURE PROPER AIR CIRCULATION. Maximum ambient temperature
for installed MST-MHT modules is 45°C. See technical manuals for RA 55H or CA 400H rack
TABLE 2-1. FUNCTIONS OF INTERNAL CONTROLS
REFERENCE
DESIGNATION
CONTROL
PURPOSE
R49
+5 V REF
This is a primary adjustment for the analog circuits.
R11
EO Zero
This control is used to adjust the output voltage of the Power Module to
zero. EO Zero calibrates the Power Module in the Voltage mode.
R51
IO Zero
This control is used to adjust the output current of the Power Module to
zero. IO Zero calibrates the Power Module in the Current mode.
R50
Current Sensing Zero
This adjustment calibrates the current monitor amplifier to zero.
R52
Full Scale Current Adjust
This adjustment calibrates the programmed full scale value.
R47
VREF
This is a full scale voltage adjustment.
R48
IREF
This is a current read back accuracy adjustment.
Содержание MST 100-2MHT
Страница 2: ......
Страница 8: ...iv MST SERIES 071320 FIGURE 1 1 MST POWER SUPPLY...
Страница 13: ...MST MHT SERIES 071320 1 5 FIGURE 1 1 MST MHT POWER SUPPLY MECHANICAL OUTLINE DRAWING...
Страница 16: ......
Страница 30: ......
Страница 39: ...MST MHT SERIES 071320 3 9 FIGURE 3 1 CALIBRATION TEST SET UP O2 S2 CS GND S1 O1...