age when the unit is in voltage mode). Start
with the most significant digit of the desired
value, then use 

T

 to highlight the next digit.

For fine adjustment press the 

ADJUST

 control

in while rotating the knob to modify the least
significant digit. 

Enter the desired value on the keypad using
the number keys. For example, to program the
BOP to 75.8V, press the following keys in
order 

75.8

 then press 

ENTER

. For fine

adjustment use 

Y

, and 

U

 to modify the least

significant digit. To correct the entry before
activation press the 

CLEAR

 key to set the

value to zero and start over. When the desired
value is displayed, press 

ENTER

. This causes

the new value to appear at the output and be
applied to the load if the output is enabled.

6. To program the corresponding Protect channel,

press 

Y

 or 

U

 as necessary to highlight the Pro-

tect channel. Then set the value using either of
the two methods described above. If the Protect
Entry setting is set to Independent, separate
entries for the positive and negative protect chan-
nel are possible. Otherwise the value entered is
applied to both positive and negative protect
channels.

NOTE: The BOP can be configured to show

the protection limits as either a single
value that applies to both protection
channels or show individual settings
for positive and negative protection
limits. See Operator’s Manual for
details

SOFTWARE LIMITS.  

Software limits prevent

programming of the main channel or the Protect
channel beyond the software limit value. Refer to
Operator’s Manual for a full explanation of software
limits.

Changing Main Channel Software limit. 

This procedure allows the user to determine the
maximum value of voltage or current that can be
programmed.

1. Press 

%

 from the power-up screen to enter the

General Setup menu, then highlight Max/Min Set-
tings. 

2. Press 

!

 to enter the Max/Min Settings submenu

(Figure 6). (If a Password is required, see Opera-
tor’s Manual for instructions.) 

3. Highlight the voltage or current max/min value

and press 

!

 to change it. Software limits are

absolute values (do not use minus sign for nega-
tive limits). Use number keys to change the set-
ting, then 

$

 to save. 

4. When complete, press 

$

 to save for power-up,

#

 to abort, or 

%

 to apply the changes (without

saving for power-up) and exit.

5. Upon return to the power-up screen, the main

channel (voltage or current) is compared against
the main channel limits in effect. If the main chan-
nel exceeds the limit, it is set to zero. 

6.  Highlight the ±CPROTECT or ±VPROTECT max/

min value and press 

!

 to change it. Software

limits are absolute values (do not use minus sign
for negative limits). Use number keys to change
the setting. Press 

$

 to save, or 

%

 to abort. 

7. When complete, press 

$

 to save for power-up,

or 

#

 to abort, or 

%

 to apply the changes (with-

out saving for power-up) and exit.

Upon return to the power-up screen, the new protec-
tion limit (voltage or current) is compared against the
protection limits in effect. If the new protection limit
setting is below the existing setting for the protection
limit, the protection channel (voltage or current) is
set to zero.

DETERMINING HOW THE UNIT RESPONDS 

WHEN OUTPUT IS OFF (LOAD TYPE) 

 The

BOP supports three Load Type selections (see Table
5) which determine how the power supply responds
when the output is off: ACTIVE, RESISTIVE and
BATTERY. The Load Type selection does not affect
the settings of the power supply for ON state; it only
affects the main internal reference level and the pro-
tection levels during the OFF state. 

WARNING

For inductive loads, and especially
superconducting magnet type
loads, the inherent offset of the
BOP in the OFF state may generate
significant current in the circuit. A
properly rated switch in parallel
with a resistor must be connected
between the power supply and the
load. The switch must be open and
the BOP front panel LCD must read
0V, 0A before removing or installing
connections between BOP and
load. 

Содержание BOP 10-100MG/ME

Страница 1: ...units the GPIB interface is replaced with a LAN interface Unless otherwise noted this guide applies to both models ACCESSING MANUALS First determine your Firmware Version see below then download the B...

Страница 2: ...or current and a 1 to 10V input for the limit channels BOP models are suitable for driving inductive loads such as large magnets or motors and for exercising batteries They are also suitable for chara...

Страница 3: ...el 1 d c Output Range Closed Loop Gain Output Impedance Voltage 2 V d c Current A d c Voltage Channel GV V V Current Channel GI A V Voltage Mode Series R L Current Mode Parallel R C Rd c mOhms L H Rd...

Страница 4: ...4 228 1692 REV 8 090418 KEPCO INC 131 38 SANFORD AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 1 BOP OUTPUT CHARACTERISTICS...

Страница 5: ...n to IEEE 1118 BITBUS port 142 0485 Kepco KMDLA 5P Kycon Inc IEEE 488 Cable 1 meter long For MG models only Connects BOP power supply to GPIB bus SNC 488 1 IEEE 488 Cable 2 meter long For MG models on...

Страница 6: ...operation voltage or current output volt age or current and establish positive and negative voltage and current limits An output analog corre sponding to output current is also provided Refer to Oper...

Страница 7: ...LS AND INDICATORS RS 232 CONNECTIONS Connect the BOP 1KW to a modem using a Null Modem patch cable at the RS 232 port located on the rear panel See Fig ure 3 A Null Modem cable is not required for old...

Страница 8: ...1355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 3 BOP 1KW SERIES REAR PANEL VIEW LINKS INSTALLED FOR LOCAL SENSING FIGURE 4 LOAD CONNECTION A LOCAL...

Страница 9: ...ng an item on the list and pressing the View Modify function key Menus and submenus will display a list of parameters with the top one high lighted The function key assignments can vary but generally...

Страница 10: ...g and output Calibration Voltage Current and External and Controls calibration Shows Firmware versions for Display Serial GPIB LAN and analog Power up Settings Mode Voltage Cur rent External Main chan...

Страница 11: ...channels one to set output voltage or current and one to set the corre sponding protection limit The main channel is deter mined by the MODE key in local mode which alternately selects either Voltage...

Страница 12: ...ral Setup menu then highlight Max Min Set tings 2 Press to enter the Max Min Settings submenu Figure 6 If a Password is required see Opera tor s Manual for instructions 3 Highlight the voltage or curr...

Страница 13: ...afety fea tures constant current type active electronic loads must be adjusted to zero and the BOP front panel LCD must read 0V minimum cur rent before handling the power supply to load connections Re...

Страница 14: ...flow during the OFF state Table 5 indicates how the power supply responds to a command to go from Output ON to OFF CONFIGURE LOAD TYPE To configure press from the power up screen then highlight Load T...

Страница 15: ...copower com email hq kepcopower com FIGURE 1 BOP 1KW OUTLINE DIMENSIONS SHEET 1 OF 2 20 000 508 00 21 439 544 54 22 800 579 11 21 564 547 73 22 000 558 79 23 863 606 12 18 235 463 16 18 985 482 21 1 4...

Страница 16: ...FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 1 BOP 1KW OUTLINE DIMENSIONS SHEET 2 OF 2 SLIDES TRAVEL DISTANCE 23 000 584 2 22 000 5...

Отзывы: