4. OPERATING PROCEDURES
The u n i t outputs the s e t values d i s p l a y e d on the v a r i o u s p o r t i o n s of the
d i s p l a y v i a the RF output t e r m i n a l . The d i s p l a y c o n s i s t s of 3 port i o n s :
MODULATION, FREQUENCY and LEVEL. How to set the various display portions i s
described below.
4-1 FREQUENCY SETTING
4-1-1 NORMAL SETTING
Enter the value by using the CURSOR key ® and rotary encoder @.
1) When the ^ ^ or • • key i s pressed, the cursor that 1 i g h t s i n the
upper portion of the d i s p l a y s h i f t s between the MODULATION, FREQUENCY
and LEVEL p o r t i o n s . Move t h e c u r s o r to t h e p o r t i o n where t h e
frequency d i g i t s are displayed.
2) Press the ^ or • key to move the cursor to the d i g i t you want to s e t .
3) When the r o t a r y encoder i s rotated, the d i g i t under the cursor s t a r t s
changing. Set to desired value.
4) Repeat steps 2) and 3) u n t i l the desired frequency i s s e t .
* I t i s not possible to enter a value that i s outside the rated frequency
range of the unit (100 kHz-llOMHz).
4-1-2 DIRECT SETTING
The set value i s entered by using the FREQ (MHz) key @, the numeric keys @,
and the ENT key © or the kHz key © .
1) When the FREQ key © i s pressed, the d i g i t s shown on the frequency
display are erased and the ENT key's LED flashes.
19