with several faults present. The switching elements in these avalanche pulsers generally have a
long life, but their life is finite and therefore it is sensible to switch off the unit or at least not
trigger it when it is not required. It is not possible to write a test routine to find 100% of all
possible faults, so if the unit appears to malfunction, and the procedures below fail to isolate the
problem please contact Kentech Instruments for advice.
There is a facility provided under the Stack Status menu to disable individual stacks. This may
be used in certain circumstances to disable faulty stacks that are causing problems. It is also
permissible to use this facility to disable output stacks to reduce the output amplitude. Note this
information is stored in non volatile memory, ie it will be remembered even when ac power is
removed. If the output amplitude appears low please check under the system status menu that
all 86 stacks are enabled.
IMPORTANT
When fitting the output connector ensure that:
i) all six fixing screws are tightened to maintain a low impedance ground connection
ii) the plastic insulating collar and sprung centre contact are fitted
Failure to fit the fixing screws will result in severe RF emission.