![Kentech Instruments PBG7 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/kentech-instruments/pbg7/pbg7_manual_1955363025.webp)
SPECIFICATIONS
1. PBG7 control and trigger module
2. Four channel delay module
3. PBG5 four channel trigger module
4. PBG7 45kV output module
5. PBG7 HV power supply module
PBG7 control and trigger module
Number of outputs
One
Polarity
Negative
Output amplitude
5.8kV into 50 ohms.
Rise time
<=100ps (10 to 90%)
Pulse width
~2ns FWHM
Maximum rep rate
>=500Hz
Trigger to output jitter
<10ps rms
Output protected against open circuit, short circuit and arcing loads.
MTBF
approx 10^10 shots
Output connector
N
Outputs:
Pulse output
N type
Pretrigger output
BNC
~10V into 50
1
, ~2ns risetime
leads main output by selectable delay when delay active
Auxiliary output
BNC
~10V into 50
1
, ~2ns risetime
approx synchronous with pulse output.
AC power
2 x IEC outlets
Splitter outputs
4 x BNC, approx 500V into 50
1
0.5ns risetime
System control
9 way D type male
Inputs:
Trigger input
BNC
5V, 50
1
, <5ns risetime
RS232
9 way D type female
9600 baud, 1 stop bit, no parity, 8 bit
pin1 = receive
pin3 = transmit
pin2 = ground
AC power
IEC, 110 to 240V ac, 50/60Hz