The cursor will be under Stack as shown.
From this cursor postion:-
Left
returns to main menu
Right
moves cursor to under 20, press
up
/
down
to change selected stack then
left
to exit.
Down
moves cursor to under Enable
With cursor under Enable
Right
moves cursor under 1
The configuration is kept in nonvolatile memory, it is stored as you exit the system status page.
Therefore if you have edited the configuration you will see this message momentarily as you
exit.
The state of the ENABLE flag is displayed. Press Down/No to return to the FAULT flag
display, Left or Right to toggle the flag, Up/Yes has no effect. The ENABLE flags are stored in
battery backed RAM.
h) REMOTE
The µP will enter REMOTE mode if it receives an ESC character while displaying the main
menu. In REMOTE, the unit accepts commands in simple ASCII characters and transmits its
responses. There is no handshake, this mode is designed to work with a simple terminal or
terminal emulator.
On entering REMOTE, the µP prints a “Kentech Instruments” banner, then waits for the user
commands. The commands are:-
RUN {cr}
to enable all stacks with enable flag set and trigger, thereafter any character will
terminate run mode and disable all stacks
TEST {cr}
to run self test routine and print out a summary of results