![Kent Scientific PhysioSuite Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/kent-scientific/physiosuite/physiosuite_user-manual_1955306009.webp)
5
To configure a new Parameter alarm:
1) From any Run Screen, touch Menu then Config.
2) Touch Data Mgmt, then Alarms.
i
Silence all Alarms
If you do not wish for the PhysioSuite to make an audible tone when an alarm
is triggered, you can create silent alarms. Turn Dial B to choose “Silence all
Alarms” or “Allow Sound”.
Touch Back when finished.
ii
Parameter 1-4
Set a custom alarm based on any measured parameter. You may choose up
to 4. Press Dial A when finished configuring the following settings:
• Parameters: Choose a parameter to monitor by alarm.
• Action: Choose how the alarm will alert you (message and/or sound)
• Type: Choose status to trigger alarm
• Value: Set the threshold value
• Delay: Choose how long the value must meet the threshold before
triggering the alarm
• Clear: Choose how the alarm needs to be cleared
iii
Fixed Alarms
The PhysioSuite includes a number of fixed alarms that alert you to scenarios
that require attention. The number of Fixed Alarms are dependent on the
number of modules installed. You may view and configure any of the Fixed
Alarms by selecting it, then pressing A to enter the Alarm Menu.
To clear a sounding alarm:
A triggered alarm produces a colored banner at the bottom of the screen describing
the cause of the alarm.
To silence the alarm:
1) Navigate to the corresponding Screen
2) Touch Setup, then Silence Alarm
OR
1) Touch Menu from any Run Screen
2) Touch Silence Alarm