The RoVent Run Screen:
RoVent Run Screen
Initial Set-Up
Ventilator Calibrations:
The first time the RoVent is used, or any time the accessories have changed, be sure to
perform a Deadspace Calibration and verify the ventilator settings.
Note:
Never perform these functions when an animal is connected to the ventilator.
1. Measure Deadspace
The Deadspace calibration will measure the volume of tubing between the
PhysioSuite RoVent module and the animal. It is important that the Deadspace
is recalibrated every time the tubing configuration is changed to maintain
accurate tidal volume delivery.
1) Set up tubing.
2) From the Vent Run Screen, touch Setup then More Ventilator.
3) Touch Calib & Tests, then Measure Deadspace.
4) Firmly seal the end of the endotracheal tube connector with your finger
or a cap, ensuring that there is no air leakage.
5) Press Dial B. When asked to confirm measurement, press Dial B again.