26
2. From any Run Screen, Touch:
1) Menu, then Data Management, then Upload
3. Choose:
1) The Parameters to upload
2) The timing
3.) Then touch Enable.
The “Always enabled” setting saves data as long as the
controller is on. The “Enabled while running” setting is
available only if your system contains a RoVent ventilator, and
saves data only when the ventilator is running.
History - Internal Storage
The History feature stores parameters records in ASCII format, up to 1000 rows. You
can send this data to a PC through the USB port at a later time. Each row can include up
to 10 Parameters simultaneously at user-selected time intervals. You can view the data
on screen or export the data to a computer after your study is complete.
Data record 1001 will overwrite data record 0001, and so on.
To save data to History:
1. From any Run Screen, Touch:
1) Menu, then Data Management, then History
2. Choose:
1) The Parameters to save
2) The timing
Tip: Changing these settings permanently deletes any data
currently saved on the system. Be sure to Export any saved
data to a PC prior to making changes.
3) Then touch Enable.