CO
2
Incubator HERAcell
®
150 Operating Instructions
59
9.
Cleaning and disinfection
CAUTION – Alcoholic disinfectants!
Disinfectants with an alcohol content of more
than 10 % may form, in combination with air,
easily combustible and explosive gas
mixtures.
When using such disinfectants, avoid open
flames or exposure to excessive heat during
the entire disinfection process!
•
Use such disinfectants only in adequately
ventilated rooms.
• After the disinfectant has been allowed to
react, wipe the cleaned device components
thoroughly dry.
• Observe safety regulations to avoid fire and/
or explosion hazard caused by alcohol-con-
taining disinfectants (ZH 1/598).
Kendro Laboratory Products recommend the disinfectant Barrycidal 36.
When applied properly, Barrycidal 36 is a highly effective broad-range disinfectant.
The effectiveness of the product is the result of its ammonium compounds. The
broad-range disinfectant is effective against viruses, bacteriae, yeasts, fungu-
ses and AIDS causatives (HIV). Barrycidal 36 is DGHM-listed.
Part no. for Barrycidal 36: 50 052 425 and 50 051 939
Restrictions:
In some European countries and in the U.S.A., Barrycidal 36 has not been ap-
proved as a disinfectant. In these areas, some other suited disinfectant that meets
the aforementioned safety requirements must be used.
As an alternative for the U.S.A., Kendro Laboratory Products recommends Micro-
cide SQ. Microcide SQ is EPA-listed.
Direct order: www.globalbio.com.
Содержание HERAcell 150
Страница 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Страница 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Страница 84: ...Internet http www kendro com...
Страница 85: ......
Страница 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Страница 106: ......
Страница 152: ......
Страница 153: ......
Страница 154: ......
Страница 155: ......
Страница 156: ......
Страница 157: ......
Страница 158: ......
Страница 159: ......
Страница 160: ......
Страница 161: ......
Страница 162: ......
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 166: ......
Страница 167: ......
Страница 168: ......
Страница 169: ......
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ......
Страница 175: ......
Страница 182: ...Assembly Instruction Lockable door for HERAcell HERAcell 150 and HERAcell 240 Edition 09 2003 50072431 A...
Страница 194: ...Installation Instructions Gas Tight Screen for the HERAcell and HERAcell 150 Retrofit Version 07 2003 50078424...
Страница 208: ...Assembly Instructions HERAcell HERAcell 150 and HERAcell 240 Main PCB Replacement Version 12 2003 50073246 D...
Страница 220: ......
Страница 221: ......
Страница 222: ......
Страница 223: ......
Страница 224: ......
Страница 225: ......
Страница 226: ......
Страница 227: ......
Страница 228: ......
Страница 229: ......
Страница 230: ......
Страница 231: ......
Страница 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Страница 249: ...Footprint HERAcell...
Страница 250: ...Footprint HERAcell...