Operating manual 

4

 

Setting parameters at the pyrometer 

6.1 

Setting the parameters using the button 

Use  the  buttons 

▲▼

 

and  the  „MODE"  button  on  the  rear  panel  to  ac-

cess  and  configure  parameters.  With  these  buttons  you  can  view  and 
adjust all settings required for pyrometer operation. 

Menu structure: 

1.

Press the MODE button while in normal operating mode to switch

to "configuration layer" mode Die Codeseite des gewünschten Pa-
rameters mit 

▲▼ 

auswählen.

2.

Use

 

▲▼ 

to select the configuration layer for the parameter you

wish to set.

3.

Press MODE to confirm. Press 

▲▼ 

to select the particular pa-

rameter.

4.

Press MODE to confirm. Press 

▲▼ 

to adjust the parameter val-

ue.

5.

Press MODE again to end. Press 

▲▼ to select END/SAVE

6.

Now use MODE to either apply the changes you have made to
the settings [

SAVE

] or to close without saving these changes

[

END

]. After that, the display panel will resume showing the tem-

perature reading.

6.2 

Operation of the pyrometer via terminal program 

By  default,  the  pyrometer  has  fully  integrated  communications  software 
to the point-to-point connection with a PC. As an alternative to the  
CellaView software, all parameters that are required for a measurement 
data acquisition or a general configuration of the pyrometer can be con-
figured via a simple terminal connection using the terminal program. 

Most  key  parameters  are  directly  available  in  the  main  menu.  Further 
functions are contained in submenus. Navigation within the menus is ex-
plained in the following graphics: 

Normal 

operating mode 

(temperature) 

Select  

config. layer 

C000- 

C999

 

Select 

parameter 

EPS 

Adjust 

parameter 

99.9 

MODE 

MODE 

MODE 

▲▼

▲▼

▲▼

Содержание PYRO CellaTemp PA

Страница 1: ...PYROMETER CellaTemp PA Mixed Mode Function Ident no 1117971 06 2022 MANUAL...

Страница 2: ...Operating manual...

Страница 3: ...Mode 2 4 Factory setting of the mixed mode function 3 5 Adjustment of the measured values 3 6 Setting parameters at the pyrometer 4 6 1 Setting the parameters using the button 4 6 2 Operation of the...

Страница 4: ...Operating manual...

Страница 5: ...start and Measuring range end 2 Mixed mode function for extending the measuring range With mixed mode operation the measuring range can be extended to lower initial temperatures by linking a spectral...

Страница 6: ...mbda 1 L2 Lambda 2 Q Quotient SRC C Selction of the con trolling source L1 Lambda 1 L2 Lambda 2 Q Quotient SAVE Save Save changes exit menu ESC Escape Discard changes exit menu To operate the pyromete...

Страница 7: ...s In order to ensure a linear transition of the measured values the adjust ment must be carried out as follows during commissioning 1 Align the pyrometer to the target object Set the two colour emissi...

Страница 8: ...VE 6 Now use MODE to either apply the changes you have made to the settings SAVE or to close without saving these changes END After that the display panel will resume showing the tem perature reading...

Страница 9: ...Show this help site J Show diagnosis W Show ambient temperature Q Show calibration data X Show measure temperatures P Show channel parameters 6 2 2 Parameter diagnostic overview Command P shows an ov...

Страница 10: ...ixed Mode you can adjust all relevant parameters In the sub menu Mixed Mode you can adjust all relevant parameters Activation deactivation of the temperature calculation Measuring range of the Mixed M...

Страница 11: ...command T in the main menu you can select the source for the analogue output Select 4 for Mixed Mode The display shows the Mixed Mode temperature T Set Analog Out 1 SOURCE 1 Lambda 1 2 Lambda 2 3 Quot...

Страница 12: ...or in part for any purpose whatsoever without prior consent of the author This applies to any form of mechanical or electronic reproduction as well as to electronic transmission in any form through a...

Отзывы: