Keller PYRO CellaTemp PA Скачать руководство пользователя страница 5

Operating manual 

1

 

General description. 

The mixed-mode function is used to link two measured values of a ratio 
pyrometer and to generate a continuous analogue output signal.  
In the lower measuring range up to the start temperature of the transition 
range, the measured value of the set source 1 is output.  
Above  the  defined  end  temperature  of  the  transition  range,  the  output 
signal  corresponds  to  the  measured  value  of  source  2.  A  sliding  transi-
tion  takes  place  between  the  start  and  end  temperature,  in  which  the 
measured values of the two sources are added together, weighted as a 
percentage. The set controlling source determines the reference value of 
the  initial  and  final  temperature  of  the  transition  range.  In  addition,  the 
entire  measuring  range  of  the  mixed-mode  channel  can  be  set  via  the 
parameters "Measuring range start" and "Measuring range end. 
 
 

 

 

 

Mixed-mode function for extending the measuring 
range 

With  mixed-mode  operation,  the  measuring  range  can  be  extended  to 
lower  initial  temperatures  by  linking  a  spectral  and  a  quotient  tempera-
ture.  At  the  beginning  of  the  measuring  range  the  device  works  as  a 
spectral  pyrometer  and  from  the  end  of  the  transition  range  as  a  ratio 
pyrometer. 
Since the pyrometer provides different measured values for the spectral 
and quotient temperature depending on the emissivity of the measuring 
object  and  the  degree  of  partial  filling  of  the  measuring  spot  (if  the 
measuring  object  is  smaller  than  the  measuring  field  of  the  pyrometer), 
the two measured values must be matched to each other in the middle of 
the transition range for a linear transition.  
For  this  purpose,  the  emissivity 

ε1  or  ε2  (depending  on  the  selected 

source) must be adjusted so that the two temperature values are identi-
cal in the middle. 
 

Temperature 

Persentage 
partition 

Controlling source 

Measuring range Mixed mode 

Transition region 

Содержание PYRO CellaTemp PA

Страница 1: ...PYROMETER CellaTemp PA Mixed Mode Function Ident no 1117971 06 2022 MANUAL...

Страница 2: ...Operating manual...

Страница 3: ...Mode 2 4 Factory setting of the mixed mode function 3 5 Adjustment of the measured values 3 6 Setting parameters at the pyrometer 4 6 1 Setting the parameters using the button 4 6 2 Operation of the...

Страница 4: ...Operating manual...

Страница 5: ...start and Measuring range end 2 Mixed mode function for extending the measuring range With mixed mode operation the measuring range can be extended to lower initial temperatures by linking a spectral...

Страница 6: ...mbda 1 L2 Lambda 2 Q Quotient SRC C Selction of the con trolling source L1 Lambda 1 L2 Lambda 2 Q Quotient SAVE Save Save changes exit menu ESC Escape Discard changes exit menu To operate the pyromete...

Страница 7: ...s In order to ensure a linear transition of the measured values the adjust ment must be carried out as follows during commissioning 1 Align the pyrometer to the target object Set the two colour emissi...

Страница 8: ...VE 6 Now use MODE to either apply the changes you have made to the settings SAVE or to close without saving these changes END After that the display panel will resume showing the tem perature reading...

Страница 9: ...Show this help site J Show diagnosis W Show ambient temperature Q Show calibration data X Show measure temperatures P Show channel parameters 6 2 2 Parameter diagnostic overview Command P shows an ov...

Страница 10: ...ixed Mode you can adjust all relevant parameters In the sub menu Mixed Mode you can adjust all relevant parameters Activation deactivation of the temperature calculation Measuring range of the Mixed M...

Страница 11: ...command T in the main menu you can select the source for the analogue output Select 4 for Mixed Mode The display shows the Mixed Mode temperature T Set Analog Out 1 SOURCE 1 Lambda 1 2 Lambda 2 3 Quot...

Страница 12: ...or in part for any purpose whatsoever without prior consent of the author This applies to any form of mechanical or electronic reproduction as well as to electronic transmission in any form through a...

Отзывы: